Product DescriptionThe MS400-DP Mirror Dew Point Meter for natural gas uses the chilled-mirror principle as a portable intelligent humidity measurement device. It integrates advanced signal detection, fuzzy-logic control and data processing, and uses high-performance imported core components to deliver accurate and reliable measurements.
Features- Intelligent operation — Stores, retrieves, prints and automatically uploads measurement records.
- Precise and fast — Built-in Condensation Accelerator (QCA) for rapid and stable dew point detection even at low humidity.
- Dynamic display — Real-time condensation indication of the mirror surface.
- Automatic adjustment — Auto-adjusts light intensity and supports automatic mirror inspection.
- Multi-gas measurement — Suitable for trace moisture analysis in a wide range of gases including SF6.
- High-performance cooling — Maximum cooling differential >95°C for stability under extreme conditions.
- Portable power — Built-in high-capacity lithium battery for over 8 hours continuous use.
- User interface — Multi-language settings and an optimized user-friendly interface.
Functional parametersMeasurement range: Dew point measured from -65°C to +20°C depending on ambient temperature (examples provided for 10°C, 20°C, 35°C; at 40°C ambient the range is -55°C to +20°C). Values are for SF6 at atmospheric pressure; higher pressure raises the dew point.
Measured parameters: dew point temperature, μL/L, equilibrium process curve, gas flow, ambient temperature and humidity.
Control interfaces: manual, serial port, and remote wireless control.
Accuracy: ≤±0.2°C. Repeatability: ≤±0.1°C. Equilibrium time: typically 3–5 minutes (depending on environment and gas dryness).
Cooling: Level-4 Peltier cooling; maximum cooling differential >95°C. Sampling flow rate: 15–60 L/H (typically 25–30 L/H). Sampling gas pressure: 10 mbar min. – 10 bar max (1 kPa – 1 MPa). Display: 320×240 true-color TFT LCD. Alarms: gas flow abnormality and mirror inspection alarms. Response time: up to 2.5°C/s. Charging input: 100–240 VAC 50/60 Hz. Host input: 20 V DC 6.0 A. Max charging power: 120 W. Ambient measurement range: -40°C to 123.8°C (±0.3°C); operating temperature 0°C to 50°C. Ambient humidity measurement: 0%–100% ±2.0%. Weight: approx. 4 kg. Size: 320 mm × 330 mm × 150 mm. Battery: built-in lithium battery >8 hours operation.
ApplicationTypical application: measurement of moisture content in SF6 and trace moisture analysis in various industrial gases.
Characteristics / Technical specifications- Model: MS400-DP
- Principle: chilled (cold) mirror dew point measurement
- Dew point range: -65°C to +20°C (depending on ambient conditions)
- Accuracy: ≤±0.2°C
- Repeatability: ≤±0.1°C
- Equilibrium time: 3–5 minutes (typical)
- Cooling: Level-4 Peltier, max differential >95°C
- Sampling flow rate: 15–60 L/H (typ. 25–30 L/H)
- Sampling pressure range: 10 mbar – 10 bar (1 kPa – 1 MPa)
- Display: 320×240 TFT color LCD
- Alarms: gas flow abnormality, mirror inspection
- Response time: up to 2.5°C/s
- Power input: 100–240 VAC 50/60 Hz; Host input: 20 V DC 6.0 A; Max charging power: 120 W
- Ambient measurement range: -40°C ~ 123.8°C (±0.3°C); Operating: 0°C ~ 50°C
- Ambient humidity measurement: 0%–100% ±2.0%
- Weight: ≈ 4 kg
- Dimensions: 320 × 330 × 150 mm