Product overviewThe PTM600-T online syngas analyzer provides simultaneous multi‑component gas measurement for industrial syngas and gasification atmospheres. It uses high‑resolution NDIR for hydrocarbons and acetylene, with optional compensated TCD for accurate H2 quantification and electrochemical/O2 sensors for percent‑level oxygen monitoring. Engineered for continuous process monitoring with configurable measurement ranges.
Key features- Simultaneous detection: CO, CO2, CH4, C2H2/CnHm (NDIR)
- Optional TCD for H2 measurement; electrochemical sensor for O2 (%VOL)
- Fast response and high repeatability for process control
- 7-inch high‑resolution color display and intuitive interface
- Multiple outputs: RS485/RS‑232 and 4‑20 mA; configurable ranges
Technical specifications- Model: PTM600-T
- Measurement technologies: NDIR (CO/CH4/C2H2), TCD (H2), ECD/electrochemical (O2)
- Measurement ranges: O2 0–25%VOL; CO 0–10%VOL; H2 0–100%VOL; CH4 0–10%VOL; C2H2 0–10,000 ppm (customizable)
- Resolution: 0.01% (CH4/H2/CO/O2); 1 ppm (C2H2)
- Precision: CO2/CH4 ≤2% FS; H2S/O2/CO/H2 ≤3% FS
- Response times: IR response 15 s; T90 (NDIR) <10 s
- Repeatability: 0.5%
- Sample flow: 0.7–1.2 L/min; inlet pressure: 2–50 kPa
- Sampling condition: sample must be free of dust, water and tar; use appropriate conditioning
- Display: 7-inch high‑resolution color
- Communications: RS485/RS‑232, 4–20 mA outputs
- Power supply: 100–240 VAC
- Operating temperature: standard −10 to 40°C (option −40 to 70°C)
- Relative humidity: <95% non‑condensing
Applications- Continuous syngas monitoring in gasification and synthesis processes
- Process gas analysis for fertilizer, chemical, cement and metallurgical industries
- Flue gas composition monitoring (e.g., CEMS) and emissions screening
- Research and laboratory gas analysis requiring multi‑component measurement