Validation logger for benchtop autoclaves
Product information
Validation logger for thermal testing of processes with steam sterilizers. The logger is characterized by its highly flexible cable sensors, which enable precise positioning. With its 3 temperature sensors and the pressure sensor it is ideally suited for measurement in small sterilizers. The very thin temperature sensors with a diameter of 1.2 mm at the sensor tip enable a very fast response time. When using the logger in the washer-disinfector, it is strongly recommended to fix the sensors by means of silicone cable ties.
Benefits:
• highly flexible cable probes
• high accuracy
• Realtime data transmission
• fast response time
• sterilizable data logger
Applications:
Validation logger for Measurement in the steam sterilizer
Technical Data
Weight
approx. 120 g
Accuracy
±0,1 °C (0 °C ... +120 °C) ±0,05 °C (+120 °C ... +140 °C) ±0,1 °C (+140 °C ... +150 °C)
±15 mbar (1 mbar ... 50 mbar), ±10 mbar (50 mbar … 150 mbar), ±10 mbar (2.050 mbar … 2.250 mbar),
±10 mbar (3.000 mbar … 3.250 mbar), ± 15 mbar (3.250 mbar ... 3.500 mbar),
± 0,5 % FS (3.500 mbar ... 4.000 mbar)
Protection class
IP68
Dimensions
ø 48 mm x 32 mm - Needle length 1.200 mm
Resolution
0,01 °C 1 mbar
Certificate
Factory calibration certificate
Measured variables
Temperature, pressure
Intervall
250 ms ... 24 h
Battery
3.6 V lithium battery exchangeable
Measuring range
Temperature 0 °C … +150 °C Pressure 1 … 4.000 mbar
Storage temperature
-20 °C ... +70 °C