Vacuum leak tester K3060
differential pressure decaydigitalpressure

Vacuum leak tester - K3060 - ForTest GmbH - differential pressure decay / digital / pressure
Vacuum leak tester - K3060 - ForTest GmbH - differential pressure decay / digital / pressure
Add to favorites
Compare this product

Characteristics

Type
vacuum, differential pressure decay, digital, pressure, leakage
Applications
for packaging, for the pharmaceutical industry, for automation
Other characteristics
in-line, compact, USB, LAN, color display, Modbus

Description

Product overview
The K3060 is a micro differential leak tester engineered for integration on high‑productivity automatic lines. Its compact differential design reduces parasitic volume and enables fast, repeatable tests on micro‑volumes.

Key features
  • Compact footprint: 110 × 110 × 110 mm
  • Differential measurement optimized for micro volumes
  • Low‑energy micro‑valves for short cycle times
  • High resolution and repeatability

Application sectors
  • Industrial automation
  • Packaging and assembly lines
  • Test benches and production inspection
  • Medical device manufacturing

Compact design for industrial automation
The unit's small dimensions allow placement close to the part under test, minimizing parasitic volume and improving cycle time efficiency for in‑line inspection.

Performance on micro volumes
Internal low‑energy micro‑valves and fluidic layout are sized for cavities up to 50 cc, enabling ultra‑fast differential tests on micro components in automated machines.

Construction and user interface
Anodized aluminium chassis with plexiglass front and integrated 3.5" colour display (320×240) for immediate status and result monitoring.

Connectivity and communication
Standard interfaces: 24 Vdc digital I/O, USB (master), RS232, RS485, optional Ethernet (ForTest or Modbus‑RTU), and Bluetooth. Operating measurement range: -1 to 10 bar.

Programming and usage methods
Programmable via dedicated PC software (LAN, USB or RS232) or USB stick. Stand‑alone operation supported using a simulated keyboard and digital I/O for Start, Abort cycle, program recall and result outputs. Up to 100 programs storable locally.

Technical specifications
  • Model: K3060
  • Operating measurement range: -1 to 10 bar
  • Intended test volume (max): up to 50 cc
  • Accuracy Δp: 0.5% RDG + 3 DGT
  • Δp measurement ranges: 0–65 mbar (lower scales); up to 0–650 mbar (highest scale)
  • Δp resolution: 0.001–0.01 mbar (depending on scale)
  • Direct pressure accuracy: 0.5% FS
  • Direct pressure resolution: 0.1 mbar (typ.) / 1 mbar (highest range)
  • Leak measurement accuracy (cc/min): 1% RDG + 0.03 cc/min
  • Leak measurement range: 0–10 cc/min; resolution 0.01 cc/min
  • Supported units: mbar, bar, psi, mmHg, mmH2O, Pa, hPa, cc/min, cc/h, pressure/s
  • Dimensions: 110 × 110 × 110 mm; Weight: 1 kg
  • Display: 3.5" (320×240)
  • Operating temperature: 5–40 °C
  • RS232: 1; RS485: 1; USB (master): 1; USB (slave): 0
  • Ethernet: optional (ForTest or Modbus RTU)
  • I/O signals (standard): Start, Stop, Filling, Test, Good, Reject, 3BCD
  • Number of programs: 100; Test archive memory: No; Password lock: No
  • Notes: Resolution for uncertainty consideration: 1 Pa. Accuracy data referenced to laboratory calibration conditions. Flowmeters may report mass flow at reference conditions (0 °C, 1013.25 mbar); leak test flow is volumetric. For higher pressure ranges accuracy is specified at offer stage. Reference norms: EN 61010‑1, EN 61326‑1, EN 61000 series, EN1779.

Catalogs

No catalogs are available for this product.

See all of ForTest GmbH‘s catalogs
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.