Product overviewWinner2006B is an intelligent, fully automated laser diffraction particle size analyzer based on Mie scattering theory. It measures particle size distributions from 0.1 μm to 1000 μm using a high-performance laser and integrated sample preparation. The instrument is suitable for research institutes, university labs, R&D centers and quality control laboratories.
Main specifications (general)- Model: Winner2006B
- Principle: Mie scattering
- Test speed: fastest 10 seconds
- Sampling frequency: up to 10 kHz
- Outer dimensions: 900 mm × 400 mm × 500 mm
- Net weight: 40 kg
- Operation modes: software operation, fully automatic (switchable), SOP one-click automatic test
Measurement size and dispersion- Measurement range: 0.1–1000 μm
- Number of channels: 76
- Accuracy error: ≤0.5% (CRM D50)
- Repeatability error: ≤0.5% (CRM D50)
- Dispersion methods: ultrasonic (40 kHz, 60 W, time adjustable), mechanical stirring (0–3000 rpm), built-in circulation (optional: 8 L/min)
- Sample tank volume: 450 ml
Optics- Light source: laser λ = 639 nm, P > 2 mW
- Optical bench alignment: automatic alignment
- Optical path: Reverse Fourier (folded optical path)
- Laser safety: Class 1
Software functions- Analysis models: free distribution, R-R distribution, log-normal and statistical models by category
- Statistical comparison: compare multiple test results across batches or processing steps
- Data processing: intelligent statistical selection of representative results
- Custom outputs: any percentile between D1 and D100; cumulative percentages above/below or within ranges
- Output formats: particle size values, differential distribution data/graphs, cumulative distribution curves; export to Word, Excel, BMP, TXT, PDF
- One-click operation: automatic water intake, air bubble removal, dispersion, measurement, cleaning and data processing
- Interface languages: Chinese and English; additional languages on request
Advantages- One-button automated workflow reduces manual handling and variability
- Wide measurement range (0.1–1000 μm) including submicron capability
- Integrated wet dispersion: ultrasonic, mechanical stirring and circulation to prevent particle re-settling
- Advanced optical design with concentrated light Fourier transform and dual-spectrum path for improved scattered light collection
- Automatic optical alignment with hybrid stepping motor for micro-positioning accuracy down to 0.1 μm
- Unconstrained free-fitting algorithm to reflect true particle distributions
Applications- Abrasives
- Adhesives
- Agrochemicals
- Barite
- Batteries
- Bentonite
- Boron carbide
- Brucite
- Calcite
- Calcium carbonate
- Carbon black
- Catalysts
- Cement
- Ceramics
- Chemicals
- Clay
- Coal
- Coatings
- Corundum
- Cosmetics
- Diamond powder
- Dolomite
- Diatomite
- Emulsions
- Environmental samples
- Explosives
- Ferrite
- Flour
- Fluorescent materials
- Fluorite
- Food & beverage
- Food additives
- Graphite
- Inks
- Kaolin
- Pharmaceuticals
- Metal powders
- Mica
- Minerals
- Oxides
- Paints
- Paper
- Petrochemical
- Pigments
- Plastics & polymers
- Quartz
- Refractories
- Resins
- Silica
- Slurries
- Soil and sediments
- Starch
- Sulfur
- Talc
- Toners
- Wollastonite
- Zeolites
- Zirconium silicate
Technical specifications- Model: Winner2006B
- Measurement range: 0.1–1000 μm
- Principle: Mie scattering
- Light source: laser λ = 639 nm, P > 2 mW
- Channels: 76
- Accuracy & repeatability: ≤0.5% (CRM D50)
- Test speed: fastest 10 s
- Sampling frequency: up to 10 kHz
- Sample tank: 450 ml
- Dispersion: ultrasonic 40 kHz/60 W, stirring 0–3000 rpm, optional circulation 8 L/min
- Optical path: Reverse Fourier, automatic alignment
- Laser safety class: Class 1
- Dimensions: 900 × 400 × 500 mm
- Weight: 40 kg
- Operation modes: software, fully automatic, SOP one-click
- Export formats: Word, Excel, BMP, TXT, PDF
- GMP compliance: optional