OverviewWinner2018 Basic is a fully automatic wet laser particle size analyzer based on Mie scattering, designed for rapid and repeatable granulometry of narrow-distribution samples. Measurement range: 0.1–450 μm. The instrument supports one-key SOP operation and is suited to research institutes, university labs, R&D centers and quality control departments.
Applications- Metallurgy: metal oxides (alumina, aluminium hydroxide, zinc oxide, iron oxide, manganese oxide, lead oxide, etc.)
- Non-minerals: calcium carbonate, talc, kaolin, diatomite, graphite, mica, nepheline, pyrophyllite
- Chemicals: pigments, conductive pastes, fillers, dyes, paints, resins, plastic powders, synthetic rubber, photosensitive materials, toner, carbon black, inks, flame retardants, fluorescent powders, foaming agents, silica, microsilica, papermaking additives
- Agrochemicals: seed coatings, suspending agents, pesticides, herbicides
- Abrasives: silicon carbide, silicon nitride, emery, corundum, micron powders
- Energy & environment: coal-water slurry, power plant desulfurization slurries, water quality and dust monitoring
Advantages- One-button fully automatic operation: SOP sequence automates water intake, degassing, dispersion, measurement, cleaning and data processing to reduce operator error.
- Integrated wet dispersion: ultrasonic (40 kHz / 60 W) + mechanical stirring (0–3000 rpm) with optional circulation (8 L/min) for consistent dispersion and reduced re-aggregation.
- Advanced optical design: Winner patented concentrated light Fourier transform and dual-spectrum optical path for wide dynamic range without lens-aperture constraints.
- Automatic optical alignment: precision hybrid stepping motor provides micro-motion centering down to 0.1 μm for stable, repeatable results.
- Automatic cuvette cleaning: ultrasonic cleaning at the sample window reduces manual maintenance and extends component life.
- Free fitting algorithm: original unconstrained fitting better reflects true particle size distributions.
Main specifications- Model: Winner2018
- Principle: Mie scattering (laser diffraction)
- Measurement range: 0.1–450 μm
- Channels: 40
- Accuracy / Repeatability: ≤0.5% (CRM D50)
- Fastest test time: 10 s
- Sampling rate: up to 10 kHz
- Light source: 639 nm laser, >2 mW (Class 1)
- Sample tank: 450 ml
- Dispersion: ultrasonic 40 kHz/60 W (adjustable), mechanical stirring 0–3000 rpm, optional circulation pump 8 L/min
- Dimensions (W×D×H): 822 × 365 × 495 mm
- Net weight: 40 kg
- Operation modes: software/manual and fully automatic one-click SOP
Software & Reporting Functions- Analysis models: free distribution, R-R, log-normal and other statistical models.
- Custom outputs: any percentile between D1–D100, cumulative percentages above/below or within ranges.
- Comparison & statistics: multi-sample/batch comparison and intelligent statistical reporting.
- Output formats: differential and cumulative distributions, particle size metrics, histograms and curves.
- Report export: Word, Excel, image (BMP), text (TXT), PDF.
- Languages: Chinese and English GUI (other languages on request).