Product descriptionThe BINDER MKT 240 is an environmental simulation chamber for cyclic temperature tests with a temperature range from -70 °C to 180 °C. It combines rapid temperature change capability with APT.line™ preheating technology to ensure homogeneous climate conditions and reliable test cycles for components and materials.
Benefits- Uniform internal climate thanks to APT.line™ preheating chamber
- Comprehensive programming and data acquisition options (touchscreen controller, internal logger, USB export)
- Large heated viewing window with LED interior lighting for sample inspection
Important characteristics- Temperature range: -70 °C to +180 °C
- 4 zero-voltage relay contacts
- Programmable condensation protection for test specimens
- Heated viewing window with LED interior lighting
- BINDER Multi Management Software APT-COM™ Basic Edition
- Troubleshooting system with visual and audible alarms
- Intuitive touchscreen controller with time-segment and real-time programming
- Internal data logger, measured values readable in open format via USB
- Access port with silicone plug: 50 mm, left
- Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
- 4 castors, two with brakes
- Computer interface: Ethernet
- Adjustable ramp function and integrated chart recorder
- Real-time clock and door heating
- 1 stainless steel rack; inner chamber made of stainless steel
- CFC-free refrigerants R-452A and R-23; cascade compressor cooling
Technical specifications- Article numbers (examples): 9020-0386; 9020-0364
- Designation: MKT240-400V (9020-0386); MKT240-480V-C (9020-0364)
- Options: Standard / with voltage and frequency converter
- Temperature range: -70...180 °C
- Temperature uniformity: 0.1...1.0 ± K (dependent on set value)
- Temperature fluctuation: 0.1...0.4 ± K (dependent on set value)
- Average heating rate (IEC 60068-3-5) -40 °C to 180 °C: 5.0 K/min
- Cooling down time 180 °C → -70 °C: 95 min; average cooling rate 180 °C → -40 °C: 4.2 K/min
- Rated voltage: 400 V (9020-0386) / 480 V (9020-0364); Power frequency: 50 Hz / 60 Hz
- Nominal power: 6.5 kW; Unit fuse: 16.0 A; Phase: 3~/N/PE (400 V) or 3~ (480 V)
- Interior volume: 228 L; Net weight (empty): 380 kg
- Load per rack: 30 kg; Permitted total load: 70 kg
- Wall clearance back/side: 300 mm / 200 mm
- Sound-pressure level: 64 dB(A) / 67 dB(A)
- Number of shelves (std./max.): 1/6
- Interior W×H×D: 735 × 700 × 443 mm; Housing W×H×D: 1,115 × 1,935 × 925 mm