Product overviewThe W510/W519‑DCL4.2 online residual chlorine analyzer uses a Germany-made CL4.2 membrane-covered amperometric 2‑electrode sensor with built-in electronics to measure pH‑dependent free residual chlorine. The sensor includes an internal temperature probe for automatic temperature compensation. Zero‑point adjustment is not required; calibration must be performed using the DPD‑1 method. The analyzer is designed for water without surfactants and requires a dedicated flow cell to ensure correct flow and pressure.
Working principleThe system comprises a controller (display/control unit) and the CL4.2 free chlorine sensor. The controller displays the disinfectant (main) measurement and, if a pH electrode is fitted, the pH value. It indicates operating status and provides analog, digital and switching outputs for integration with process control systems.
Signal processing- Measurement system: membrane‑covered amperometric 2‑electrode sensor with integrated electronics.
- Output options: analog (4–20 mA), digital (Modbus RTU) and relay/switching signals.
- Built‑in temperature sensor enables automatic temperature compensation (avoid sudden temperature changes).
- No zero‑point adjustment required; calibration performed with the DPD‑1 method.
Application- Suitable for water without surfactants (tensides), since surfactants can contaminate the membrane and degrade accuracy.
- Typical water types: drinking water, swimming pools, service water and process water (must use dedicated flow cell with required flow/pressure).
Technical parametersMeasuring parameter: Free chlorine (pH‑dependent). Response time T90 ≈ 30 s. Operating conditions: water temperature 0–45 °C (no condensation), ambient temperature 0–55 °C, pH range 6–8. Required flow in flow cell: 15–30 L/h. Connection: 2‑pin terminal (2×1 mm²) or 5‑pin M12 plug. Interfering substances: chlorine dioxide (ClO2, factor ~9), ozone (O3); electrolytically produced chlorine from non‑membrane cells can cause errors.
Features / technical specifications- Sensor: Germany‑made CL4.2 free chlorine sensor, membrane‑covered amperometric 2‑electrode system with built‑in electronics.
- Main parameter: free residual chlorine (pH‑dependent).
- Calibration: DPD‑1 method required.
- Temperature compensation: internal temperature sensor, automatic compensation (avoid rapid temperature changes).
- Output options: 4–20 mA, Modbus RTU, switching signals.
- Measurement ranges (examples): Modbus RTU 0.005–2.000 ppm (0.01 mg/L resolution), 0.5–20.0 ppm (0.1 mg/L); 4–20 mA selectable ranges include 0.005–0.500 ppm, 0.005–2.000 ppm, 0.05–5.00 ppm, 0.05–10.00 ppm, 0.5–20.0 ppm (resolutions as above).
- Response time: T90 ≈ 30 s.
- Operating temperatures: water 0–45 °C (no condensation); ambient 0–55 °C.
- pH range: 6–8.
- Required flow (flow cell): 15–30 L/h.
- Connections: 2‑pin terminal (2×1 mm²) or 5‑pin M12 plug.
- Limitations: not suitable for water containing surfactants; affected by O3 and ClO2; electrolysis without membrane may cause measurement errors.