OverviewThe THR Series (Temperature Humidity Room) is a walk-in chamber engineered for large parts and machinery requiring combined temperature and humidity control. The room integrates pipe and duct systems, an all-in-one touch-screen control panel and a programmable logic controller. Internal dimensions, temperature and humidity ranges, and special options are customizable to suit DUT size and laboratory layout.
Key Capabilities- Precise low-humidity control down to 5% R.H. (special option with dehumidifier).
- Six selectable low-temperature grades (A–F) down to −70 °C; standard high-temperature up to 80 °C (80–150 °C optional).
- PID control for temperature and humidity with real-time humidity sensor and USB data storage for test export.
- Programmable profiles: 100 patterns × 999 segments each, supporting ramp-rate control.
- 50 mm cable test port for routing cables to the DUT during test cycles.
- Double-glazed heated observation window with interior lighting.
SpecificationsParameter: Specification
Model: THR-volume — low-temperature options A~F
Internal Dimensions: Factory direct — internal dimensions customizable
Temperature Range: Low: (A) 25 °C / (B) 0 °C / (C) −20 °C / (D) −40 °C / (E) −60 °C / (F) −70 °C
High: 80 °C (80 °C ~ 150 °C optional)
Humidity Range: 20% ~ 98% R.H. (10%–98% R.H. / 5%–98% R.H. available as special option with dehumidifier)
Indication Resolution / Distribution Uniformity: 0.1 °C; 0.1% R.H. / ±2.5 °C; ±3.5% R.H.
Control Accuracy: ±0.5 °C; ±2.5% R.H.
Temperature Rising / Falling Velocity: Rising approx. 0.1 ~ 3.0 °C/min; Falling approx. 0.3 ~ 1.0 °C/min (faster change rate optional)
Cooling System: Water-cooled or air-cooled
Compressor: Tecumseh (France); Bizer (Germany)
Protection Devices: Fuse-free switch; compressor overload protection; high/low coolant voltage protection; over-humidity & over-temperature protection; fuses; fault warning system; water short-storage warning protection
Optional Accessories: Slope for cart; recorder; water purifier; dehumidifier
Power: AC 380 V (±10%), 50/60 Hz, 3-phase 5-wire
Features & Safety- Seven independent protection layers covering power, motor, compressor, heater and water supply to reduce risk to samples and operators.
- Heated, sweat-proof viewing window for clear observation under humid conditions.
- External water supply with convenient refill to the humidifying pot; auto-recyclable.
- Automatic refrigerant demand control to improve energy efficiency compared with conventional systems.
Standards & Applications- Supported test standards: IEC 68-2-1, IEC 68-2-2, IEC 68-2-3, IEC 68-2-30, IEC 68-2-14, MIL-STD-810G/F, ISO 16750-3.
- Typical applications: full-product temperature and humidity testing for computers & monitors, printers & fax machines, LED & PCB assemblies, automotive components, home appliances, electronics & telecommunications, research & university labs, testing & certification bodies.
On-site Installation & Delivery- On-site installation and commissioning available; services include site layout and utility planning, chamber assembly and connection, temperature and humidity calibration, operator training and handover.
- Deliverables: installation & operation manual, export packing list & shipping documents, 12-month warranty covering chamber, refrigeration system and controller, lifetime remote technical support. Typical lead time for customized walk-in chambers: 30–40 working days.
DescriptionWalk-in temperature and humidity chamber room for testing large samples and assemblies. Standard built-in functions include PID control, programmable profiles, humidity sensor for closed-loop control, overheat thermostat, USB data storage, cable test port and double-glazed observation window.
Technical Specifications- Series: THR Series
- Model designation: THR-volume (low-temperature options: A–F)
- Inner dimensions: customizable (factory direct)
- Temperature range: Low options 25 °C to −70 °C (A–F); High standard 80 °C (option up to 150 °C)
- Humidity range: 20%–98% R.H. (optional 10%–98% / 5%–98% with dehumidifier)
- Resolution / Uniformity: 0.1 °C; 0.1% R.H. / ±2.5 °C; ±3.5% R.H.
- Control accuracy: ±0.5 °C; ±2.5% R.H.
- Temperature change rate: Rise 0.1–3.0 °C/min; Fall 0.3–1.0 °C/min (faster optional)
- Cooling: Water-cooled or air-cooled; compressors by Tecumseh (France) or Bizer (Germany)
- Power supply: AC 380 V (±10%), 50/60 Hz, 3-phase 5-wire
- Protection: Multi-layer protection including overload, over-temperature, over-humidity, voltage protections and water shortage warning