IntroductionThe N530A 2.0 gas permeability tester uses the differential pressure method to measure gas transmission through films, sheets, paper and packaging materials. Designed for industrial R&D and quality control, it meets major international standards and supports a wide range of gases and barrier levels.
Product profileBuilt in accordance with GB/T 1038 and ISO 2556, the N530A 2.0 integrates high-precision pressure sensors and automatic pressure-difference maintenance. The instrument provides reliable, repeatable measurement across high, medium and low barrier materials for sectors such as food, pharmaceuticals, medical devices, consumer goods and electronics.
Test principleThe instrument operates on the differential pressure principle: the sample is mounted between upper and lower chambers, the system evacuates the gas path, the upper chamber is pressurized with test gas, and the pressure change in the lower chamber is recorded to calculate transmission rate and barrier properties.
Compliance standards- ChP 2025 4007
- GB/T 1038.1
- ISO 2556
- ISO 15105-1
- ASTM D1434
- JIS K7126-1
- YBB 00082003
Technical parameters- Test range: 0.02–50000 cm3/(m2·24h·0.1MPa), expandable to 600000
- Resolution: 0.0001 cm3/(m2·24h·0.1MPa)
- Temperature control: 15–60 °C, ±0.1 °C
- Vacuum degree: <10 Pa, 0.01 Pa resolution
- Test pressure: standard 0.1 MPa (0–300 kPa optional)
- Test gases: O2, CO2, N2, Air, He, etc.
- Test chambers: 3 independent channels
- Dimensions: 690×620×390 mm
- Weight & power: 55 kg, 750 W, AC 220V 50Hz
Product features- High-precision measurement with automatic pressure-difference control and 0.01 Pa resolution
- Three independent test chambers to run identical or different samples simultaneously for higher throughput
- Comprehensive permeation analysis: transmission rate, permeability, diffusion and solubility coefficients with data fitting
- Real-time visual curves for transmission, temperature and pressure
- GMP-compliant software: audit trail, multi-level user rights and electronic signature; report export
- Nationally certified calibration for traceability and measurement authority
- Optional IoT platform for remote data management, diagnostics and document/video access
- Adaptable to films, sheets, bottles, bags and various packaging forms
Applications- Plastic films: PET, PE, BOPP, CPP, composite and biodegradable films
- Sheets: pharmaceutical blister sheets, PVC, PP, rubber sheets
- Paper & paperboard: coated paper, aluminum foil paper, composite board
- Medical patches and disposables: plasters, heating patches, sanitary products
- Packaging: bottles, bags, cans, cups, plastic and glass containers
- Test gases: O2, CO2, N2, air, He and other gases
Technical specifications- Model: N530A 2.0
- Test range: 0.02–50000 cm3/(m2·24h·0.1MPa), extendable to 600000
- Resolution: 0.0001 cm3/(m2·24h·0.1MPa)
- Temperature control: 15–60 °C, ±0.1 °C
- Vacuum degree: <10 Pa, 0.01 Pa resolution
- Test pressure: standard 0.1 MPa (optional 0–300 kPa)
- Supported gases: O2, CO2, N2, Air, He
- Chambers: 3 independent channels
- Dimensions: 690×620×390 mm
- Weight & power: 55 kg, 750 W, AC 220V 50Hz