Product overviewFlexible strip-type biosensor designed for dip-in applications and continuous in-line lactate monitoring. Single-analyte configuration (lactate as standard) with on-chip reference, counter and blank electrodes for reliable background compensation and stable measurements.
Sterilization- Recommended dose: < 25 kGy
- Sensitivity increases with applied irradiation dose
- Lifespan decreases with applied irradiation dose
- Initial bioburden: < 1 CFU per sensor
- Do not use organic solvents
Benefits & Characteristics of the IV4 biosensor- Long-term operational stability suitable for continuous monitoring
- Extended shelf life under recommended storage conditions
- Compatible with gamma and beta irradiation sterilization
- Fast response time and high measurement repeatability
- On-chip reference, counter and blank electrodes
Usage limitation / DisclaimerEvaluation product for professionals to be used solely for research and development purposes. Not for medical or diagnostic use. Not to be used on humans.
QualityIST AG is certified according to ISO 9001:2015 (quality) and ISO 14001:2015 (environment). Appropriate processes are part of daily work and are regularly audited and extended.
Technical specifications- Product Name: B.IV4.L.1X.JST
- Product code: 105191
- Product Old code: 390.00340
- Analyte: Lactate
- Sensor dimensions: 58.8/4 x 2.5/0.6 x 0.2 mm (L/L1 x W/W1 x H)
- Construction type: strip-type with 4 electrodes: 1x lactate, 1x blank, 1x reference, 1x counter
- Working electrode: Platinum covered with enzyme membrane
- Blank electrode: For background compensation
- Reference electrode: Silver / silver chloride
- Counter electrode: Platinum
- Measurement range: 0.1 mM to 50 mM
- Sterilization: Irradiation (beta, gamma); recommended dose < 25 kGy
- Shelf life: > 6 months (at recommended storage conditions)
- Operational life time: > 4 weeks in analyzer mode