High-temperature calibration bath Calib’Bath
low-temperaturehigh-precisionfor temperature sensors

High-temperature calibration bath - Calib’Bath - JRI - low-temperature / high-precision / for temperature sensors
High-temperature calibration bath - Calib’Bath - JRI - low-temperature / high-precision / for temperature sensors
High-temperature calibration bath - Calib’Bath - JRI - low-temperature / high-precision / for temperature sensors - image - 2
High-temperature calibration bath - Calib’Bath - JRI - low-temperature / high-precision / for temperature sensors - image - 3
High-temperature calibration bath - Calib’Bath - JRI - low-temperature / high-precision / for temperature sensors - image - 4
Add to favorites
Compare this product

Characteristics

Options
high-temperature, low-temperature
Other characteristics
high-precision, for temperature sensors, portable
Temperature

Max.: 100 °C
(212 °F)

Min.: -20 °C
(-4 °F)

Description

Overview
Calib’Bath is a dry JRI calibration bath providing a stable and accurate temperature environment for metrological operations across a range of −20°C to +100°C. It is compatible with the JRI MySirius self-calibration procedure for automated quality calibrations and designed for probe calibration in laboratory and field settings.

Key benefits
  • High homogeneity and temperature stability for reliable calibrations
  • Ready-to-use, user-friendly design for probe calibration tasks
  • Multiple inserts available to extend calibration capabilities

Features summary
  • Precision: homogeneity and stability appropriate for metrological use
  • Ease of use: ready-to-use bath suited for calibrations of remote probes
  • Multiple inserts: standard and optional inserts for digital probes increase flexibility

Description
The Calib’Bath is a dry calibration bath that maintains a controlled and accurate temperature range for temperature probe calibration. It can perform automated quality calibrations when used with the MySirius self-calibration procedure. The standard insert enables automatic calibration of up to 8 T2 and T3 probes, including a well for a reference probe. An optional insert is available for JRI digital probes (Evolution, X², Green) and supports calibration and adjustment of up to 5 digital probes.

Advantages
  • Simple operation and quick setup
  • Stable and reproducible temperature environment for metrological tasks
  • Flexible insert options to cover a range of probe types

Technical specifications
  • Dimensions: 315 x 180 x 270 mm
  • Weight: ~12 kg
  • Interface: LCD display
  • Measurement range (at 23 °C): −20 °C to +100 °C
  • Accuracy: ±0.1 °C
  • Stability: ±0.02 °C
  • Heating time: 25 min (from 25 °C to 140 °C)
  • Cooling time: 25 min (from 25 °C to −25 °C)
  • Power supply: 115–230 Vac / 50–60 Hz
  • Inserts available: standard dry insert for T2/T3 sensors (included); dry insert for Digital Sensor, Digital Sensor X2 and Green Digital Sensor probes (optional)
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.