OverviewThe CellaTemp® PA 10 is a digital pyrometer for non-contact temperature measurement from 0 to 1,000 °C. Its wide measuring span and interchangeable optics allow use in varied industrial applications on primarily non-metallic measuring objects. Options for through-the-lens sighting, laser pilot light or an integrated video camera simplify accurate target alignment. A rear 4‑digit red LED display is readable from a distance and shows measured temperature or modified parameter values when settings are changed via rear buttons.
Key features- Measuring range: 0 – 1,000 °C
- Thermopile sensor, spectral range 8 – 14 µm
- Focusable interchangeable precision lenses (PZ series)
- High distance ratio variants: AF 1 ≈ 50:1, AF 2 ≈ 48:1
- 2 × analogue outputs (0(4)–20 mA), digital I/O, USB and RS‑485 interfaces
- Stainless steel housing, protection class IP65
Versions & optics- Versions: CellaTemp PA 10 AF 1 (focus 0.3 m – ∞) and AF 2 (focus 0.15 – 0.3 m)
- Standard lens: PZ 10.01 (0.3 m – ∞)
- Close-up lens: PZ 10.05 (0.15 m – 0.3 m)
- Sighting: through-the-lens sighting, laser pilot light, video camera (optional)
Measurement parameters (selected)- Response time t98: ≤ 30 ms
- Measurement uncertainty: 1 % of reading [°C], min. 2 K
- Repeatability: 1 K
- Resolution display: 0.1 K (< 200 °C) / 1 K (≥ 200 °C)
- Resolution interface: 0.1 K
- Resolution current output: 0.2 K + 0.03 % of selected range
Optical / measuring field- Measuring field shape: round
- Examples (distance → spot diameter): 0.00 mm → 24.00 mm; 12.00 mm → 23.28 mm; 24.00 mm → 22.56 mm; ... ; 300.00 mm → 6.00 mm
- Distance ratios: AF 1 = 50 : 1, AF 2 = 48 : 1
Electrical & interfaces- Analogue outputs: 2 × 0(4)–20 mA (linear, switchable, scalable), load ≤ 500 Ω
- Analogue input: 0 – 10 V
- Switching outputs: 2 × Open Collector 24 V DC, ≤ 30 mA
- Switching inputs: 2 × up to 24 V
- Interfaces: USB, RS‑485
- Power supply: 24 V DC (+10 % / −20 %), current consumption ≤ 175 mA
Camera (optional)- Video output (camera model): Composite PAL, 1 Vpp, 75 Ω
- Camera aids visual alignment of the measuring spot
Display, controls & functions- Rear 4‑digit red LED display (digit height 8 mm), 2 status LEDs
- Control: 3 rear buttons; display shows current value or changed parameters
- Overtemperature signal: if internal temperature > 80 °C analogue output > 20.5 mA
- Service function: simulation of measured value or current signal
- ATD (Automatic Temperature Detection) for discontinuous processes (optional)
General parameters- Connection: M12 thread 8‑pin connector
- Ambient temperature (operation): 0 – 65 °C; storage: −20 – +80 °C
- Permissible humidity: ≤ 95 % r.H. (non‑condensing)
- Dimensions: Ø 65 mm × 220 mm (max. 277 mm)
- Housing: stainless steel V2A (1.4305)
- Protection class: IP65 (DIN 40050)
- Weight: approx. 0.9 kg
Adjustable parameters- Measuring range and scaling of analogue I/O
- Smoothing filter; memory mode and hold time
- Emissivity and transmittance; compensation of reflected ambient radiation
- Alarm limits and operating mode; linearization
- Simulation of current/temperature; temperature unit °C/°F
- Camera models: TBC function, white balance
Scope of delivery- Pyrometer CellaTemp PA 10
- Connection cable VK 02/A (5 m)
- USB cable VK 11/D (1.8 m)
- Video cable VK 02/F (5 m) for camera models
- Software CellaView; short manual
Technical summary- Model: CellaTemp® PA 10
- Measuring range: 0 – 1,000 °C
- Sensor: Thermopile; spectral range 8 – 14 µm
- Response time: ≤ 30 ms (t98)
- Accuracy: 1 % of reading, min. 2 K; repeatability 1 K
- Analogue outputs: 2 × 0(4)–20 mA; interfaces USB, RS‑485
- Protection: IP65; housing stainless steel; weight ~0.9 kg