Product descriptionThe CellaTemp® PA 30 is a digital single‑colour pyrometer for non‑contact temperature measurement from 400 to 2,500 °C. Its short spectral range (0.78–1.06 µm) provides increased accuracy on metallic targets. Sighting options include through‑the‑lens aiming, laser pilot light or an optional video camera. Emissivity, measuring range, smoothing and other parameters are configured via rear buttons and displayed on the 4‑digit LED; normal operation shows the current measured temperature.
Special features- Measuring range 400 to 2,500 °C
- Focusable interchangeable lenses for precise distance adjustment
- Broadband anti‑reflective precision optics
- 4‑digit red LED display (8 mm digits) readable at distance
- Test current output for diagnostics
- Continuous light sensors — no mechanically moving parts
- Standard connectivity: digital I/O, two analogue outputs, USB and RS‑485
Available versions- CellaTemp PA 30 AF 1
- CellaTemp PA 30 AF 2
- CellaTemp PA 30 AF 3
- CellaTemp PA 30 AF 4
- CellaTemp PA 30 AF 5
- CellaTemp PA 30 AF 10
Optical variants / Lenses- Standard lens PZ 20.01 (focusing range depends on version)
- Close‑up lens PZ 20.03 (close focusing)
- Wide‑angle lens PZ 20.05
- Telephoto lens I PZ 20.06
- Telephoto lens II PA 20.06
- Objective F50 PZ 20.08 (for lower measuring ranges)
Measuring principle & fieldOne‑colour (single‑colour) pyrometer; measuring field shape: round.
Interfaces & outputs- Analogue output: 0(4)–20 mA linear, switchable, scalable (≤ 500 Ω)
- Analog input: 0–10 V
- Switching outputs: 2 × Open Collector 24 V DC, ≤ 30 mA
- Switching inputs: 2 × up to 24 V
- Interfaces: USB and RS‑485 (IO‑Link features on applicable models/options)
Camera (for models with video camera)Composite PAL video output, 1 Vpp, 75 Ω. Video camera available as a sighting option.
Scope of delivery- Pyrometer CellaTemp PA
- Connection cable VK 02/A (5 m)
- Video cable VK 02/F (5 m) for models with built‑in camera
- USB cable VK 11/D (1.8 m)
- Software CellaView
- Short manual
Measuring field tableMeasuring distance | Measuring field diameter
0.00 mm | 18.00 mm
16.00 mm | 17.36 mm
32.00 mm | 16.71 mm
48.00 mm | 16.07 mm
64.00 mm | 15.42 mm
80.00 mm | 14.78 mm
96.00 mm | 14.14 mm
112.00 mm | 13.49 mm
128.00 mm | 12.85 mm
144.00 mm | 12.21 mm
160.00 mm | 11.56 mm
176.00 mm | 10.92 mm
192.00 mm | 10.27 mm
208.00 mm | 9.63 mm
224.00 mm | 8.99 mm
240.00 mm | 8.34 mm
256.00 mm | 7.70 mm
272.00 mm | 7.06 mm
288.00 mm | 6.41 mm
304.00 mm | 5.77 mm
320.00 mm | 5.12 mm
336.00 mm | 4.48 mm
352.00 mm | 3.84 mm
368.00 mm | 3.19 mm
384.00 mm | 2.55 mm
400.00 mm | 1.90 mm
Note on configurationSelectable sighting options: through‑the‑lens sighting, laser pilot light, video camera. Multiple AF versions differ by focus distance, distance ratio and objective; version selection affects other settings.
Technical specifications- Measuring range: 400–2500 °C (partial ranges available)
- Sensor: photodiode
- Spectral range: 0.78–1.06 µm
- Response time t98: ≤ 50 ms (for mid ranges), down to ≤ 2 ms at higher temperatures depending on range
- Measurement uncertainty: 0.3 % of reading, min. 4 K
- Repeatability: 1 K
- Distance ratio (examples): AF 1 = 210:1, AF 2 = 200:1, AF 3 = 310:1, AF 4 = 55:1, AF 5 = 430:1, AF 10 = 60:1
- Power supply: 24 V DC (+10 % / −20 %), current ≤ 175 mA
- Display: 4‑digit red LED (8 mm), status LEDs, 3 control buttons
- Ambient temperature (operating): 0–65 °C; storage: −20–+80 °C
- Protection class: IP65
- Housing: stainless steel V2A (1.4305); dimensions Ø 65 mm × 220 mm (max. 277 mm); weight ≈ 0.9 kg
- Adjustable parameters: measuring range, analogue I/O scaling, smoothing filter, emissivity, transmittance, reflected radiation compensation, alarm limits, ATD parameters (optional), linearization, °C/°F
- Functions: overtemperature signalling, service simulation, ATD for discontinuous processes