Product SummaryThe RP2060 is a power/voltage rail probe designed for high-sensitivity oscilloscope measurements on DC power rails. It provides 2 GHz bandwidth, 220 ps risetime, nominal 1.2x attenuation and a wide ±60 V offset range while maintaining a ±800 mV input dynamic range.
Overview (comparison)- Bandwidth: RP2060 — 2 GHz | RP4060 — 4 GHz
- Risetime: RP2060 — 220 ps | RP4060 — 110 ps
- Offset range: RP2060 — ±60 V | RP4060 — ±60 V
- Input dynamic range: RP2060 — ±800 mV | RP4060 — ±800 mV
- Attenuation: RP2060 — 1.2x | RP4060 — 1.2x
- Non-Destruct Voltage: RP2060 — ±100 V | RP4060 — ±100 V
- DC Input Resistance: RP2060 — 50 kΩ | RP4060 — 50 kΩ
Key Features- 2 GHz bandwidth
- 220 ps risetime
- ±60 V offset range
- ±800 mV input dynamic range
- 1.2x attenuation
- ±100 V non-destruct voltage
Large Offset RangeThe ±60 V offset allows the DC level of a power rail to be placed near the vertical center of the oscilloscope display at high-sensitivity gain settings, enabling accurate observation of small signal variations within the ±800 mV dynamic range.
Low Attenuation and NoiseNominal 1.2x attenuation, combined with a DC 50 Ω coupling to the oscilloscope, minimizes additive noise and is especially effective with HDO oscilloscopes to achieve lowest noise at high sensitivity.
High DC Input ImpedanceA nominal 50 kΩ DC input resistance reduces probe loading on power rails, improving measurement accuracy and preserving signal fidelity.
Wide Assortment of Tips and LeadsThe probe ships with solder-in leads and coaxial cables featuring MCX and U.FL PCB receptacle mounts. A browser tip is available separately. Additional receptacles and leads can be purchased as accessories and left connected in-circuit for quick access to different signals during testing and validation.
Optional Browser TipAn optional 500 MHz browser tip provides a 50 Ω tip impedance with 1x attenuation via a pass-through 0 Ω resistor. The accessory includes 450 Ω and 950 Ω resistors to configure the tip as a conventional 10x or 20x transmission-line probe.
Technical Specifications- Offset Range: ±60 V
- Noise: ~110 µV additive to oscilloscope noise (when used with HD4096 oscilloscope)
- Rise Time (10–90%): 220 ps (typical, MCX receptacle or solder-in lead)
- Bandwidth: 2 GHz (guaranteed, MCX receptacle); 2 GHz (typical, solder-in lead); 2 GHz (typical, U.FL cable + receptacle); 500 MHz (typical, browser)
- Input Capacitance: 0.1 µF (in series with 50 Ω)
- Attenuation: 1.2x
- Input Dynamic Range: ±800 mV
- Non-Destruct Voltage: ±100 V
- DC Input Resistance (nominal): 50 kΩ