H2scan’s Series 700 in-line hydrogen-specific process monitors are designed for ease of use, interface flexibility and true process control. The HY-OPTIMA™ 700 is a solid state sensor that can operate in humidity and process gas streams with temperatures up to 100ºC. The HY-OPTIMA™ 700 is ideal for hydrogen production and petrochemical applications where real time measurements of hydrogen can increase process plant efficiencies, improve diagnostics and enhance maintenance management translating into higher profitability.
Performance
Hydrogen Sensitivity Range:
0.5% to 100% hydrogen by volume at 1ATM
Typical Response Time: T90 less than 30 sec
Ingress Protection: IP64 capable
Calibration Interval: 90 days
Product Life Expectancy: 10 years
Accuracy: ± (3.0% of indication + 2000ppm)
Interface Options
Input Voltage Range: 8VDC to 13VDC
Input Current: 0.5A
Analog Outputs: 4-20mA
Serial Communication Options: RS232 or RS422
Relay Contacts: Two programmable 60VDC/1A
SPDT relays with both normally open (N.O.)
and normally closed (N.C.) contacts.
Operating Conditions
Operating Humidity: 0% to 95% RH
Flow Rate: 0.1slpm to 50slpm
Pressure: 0 to 7bar gauge, 0 to 100psi gauge
Process Gas Temperature Range: -20ºC to 100ºC
Operating Temperature Range: -20ºC to 40ºC
Storage Temperature Range: -40ºC to 50ºC
Calibration Background Gas: Nitrogen
|
|
|
The analyser from ARELCO A.R.C. model CATARC MP - R integrates a sensor which works on the principle of thermal conductibility and a very powerful electronics with microprocessor. It is designed for monitoring and checking binary gas blending like H2/N2, He/N2, SF6/air, ...
|
|
H-Sense
Hydrogen Gas Analyzer
One of industry's first, true, on-line hydrogen gas analyzers. By utilizing standard MS techniques, coupled with a unique gas handling system, the H-Sense is proven to be a powerful, high speed, direct emission, hydrogen gas analyzer.
The mobile analyzer boasts a number of advantages:
Electron Pulse Ionization (EIMS) mass spec principle optimized for H2 only analysis.
Fully integrated gas inlet system compensates for high pressure variations.
High dynamic detection range (low ppm to 100%), highest degree of flexibility, ruggedized for any industrial or laboratory process.
System contamination from particulate or condensation is eliminated
|
|
The analyzers of line 4400 are suitable to being installed in rack cabinets 19 " or to panel, being placed on a job plan, and at last easy transportable thanks to a comfortable balancing handle.
They integrate the sensitive units and of control of the analyzers from process in an elegant aluminum structure and supply in continuous, with extreme precision, marks them proporziona them to variable the gaseous one wished.
|
|
| ... products without technical information |
| |
|
AMETEK Western Research® manufactures a range of H2S analyzers for applications including gas sweetening, pipeline quality/custody transfer, synthetic natural gas, blending stations, lean/rich amine and contactor overhead, tail gas treating, and any other process where % or PPM concentrations of H2S must be measured.
|
|
| ... products without technical information |
| |
|
|