1. Products
  2. BERG Kompressoren GmbH

Oil vapor sensor METPOINT®️

Oil vapor sensor - METPOINT®️ - BERG Kompressoren GmbH
Oil vapor sensor - METPOINT®️ - BERG Kompressoren GmbH
Oil vapor sensor - METPOINT®️ - BERG Kompressoren GmbH - image - 2
Oil vapor sensor - METPOINT®️ - BERG Kompressoren GmbH - image - 3
Oil vapor sensor - METPOINT®️ - BERG Kompressoren GmbH - image - 4
Oil vapor sensor - METPOINT®️ - BERG Kompressoren GmbH - image - 5
Add to favorites
Compare this product

Description

Product information
METPOINT®️ OCV compact — oil vapour measuring instrument for compressed air systems. Continuous on‑line monitoring detects hydrocarbon vapours and gases with a detection limit down to 0.001 mg/m³. Short measurement intervals provide fast indication of small deviations, enabling documentation of compressed air quality and identification of contamination sources.

Function of oil vapour measuring
The OCV compact uses a PID photo‑ionisation detector (UV lamp). Measured values are temperature‑ and pressure‑compensated to meet ISO 8573 requirements. Continuous self‑control is achieved by alternating measurement air with a generated reference gas, ensuring reproducible and reliable readings.

Internal reference gas generation
Integrated reference gas generation
An integrated patented catalytic converter generates reference gas by decomposing hydrocarbons to H2O and CO2 on a heated catalyst surface. This reference gas is routed alternately with the compressed air sample into the measuring chamber to clean the chamber and check the zero point for long‑term accuracy.

Technical data (extract)
Medium: compressed air (free of aggressive, corrosive, toxic, flammable or oxidising substances); upstream treatment required.
Measured value: residual oil content in mg oil/normal m³ (1.0 bar, +20 °C, 0 % RH) according to ISO 8573‑1.
Detectable substances: polyalphaolefins, aromatic and aliphatic hydrocarbons, functional hydrocarbons.
Measuring range: ≤ 0.01 … 2.50 mg/m³; detection limit 0.001 mg/m³. New measured value every 4 seconds. Calibrated range ≤ 0.01 … 1.25 mg/m³ (ISO 8573‑1).

Characteristics / technical specifications
  • Model: METPOINT®️ OCV compact
  • Measuring range: 0.01 … 2.50 mg/m³ (detection limit 0.001 mg/m³)
  • Calibrated range: 0.01 … 1.25 mg/m³ (ISO 8573‑1)
  • Measurement update: every 4 seconds
  • Accuracy: ≤0.01–0.5 mg/m³ ±0.003; ≥0.5–1.0 mg/m³ ±0.10; ≥1.0–2.5 mg/m³ ±0.10
  • Compressed air connection: G 1/8” (ISO 228‑1)
  • Operating pressure: 3–16 bar(g) (optional reducer up to 300 bar(g))
  • Power supply: 100–240 VAC, 50–60 Hz
  • Interfaces: 4–20 mA, RS‑485 MODBUS RTU, 1 alarm contact
  • Degree of protection: IP54
  • Dimensions (W x H x D): 410 x 440 x 163 mm
  • Weight: approx. 16.3 kg
  • Reference gas: integrated catalytic converter for internal reference gas generation and self‑monitoring

Catalogs

No catalogs are available for this product.

See all of BERG Kompressoren GmbH‘s catalogs
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.