Product descriptionThe KBF-S ECO 1020 is a constant climate chamber using thermoelectric (Peltier) cooling with patented heat dissipation. The design delivers low energy consumption and quiet operation and is suitable for regulated stability and environmental tests (ICH Q1A compliance possible with appropriate configurations).
Benefits- Safe: supports standard-compliant testing according to ICH Q1A, including under full load conditions.
- Reliable: engineered for failsafe operation and comprehensive self-test diagnostics.
- Adaptable: extensive accessory range enables customer-specific configurations and integrations.
- Economical: optimized thermoelectric cooling minimizes energy use and avoids refrigerants.
Important characteristics- Temperature range: 0 °C to +70 °C (max. 24 °C below ambient for full range).
- Humidity range: 10 % to 80 % RH (combined control: 5...70 °C).
- APT.line™ preheating chamber technology; independent water supply via tank.
- LCD display for temperature, humidity, alarms and additional information.
- Internal data logger; measured values exportable in open format via USB; Ethernet interface for data and alarms.
- Unit self-test for comprehensive status analysis; inner chamber fully stainless steel.
- Standard: 2 stainless-steel racks (up to 15 shelves max.); access port with silicone plug Ø 30 mm (left); sound pressure level ~49 dB(A).
Technical specifications- Article Numbers: 9020-0419 (KBFSECO1020-230V) / 9020-0422 (KBFSECO1020UL-120V).
- Temperature range (with humidity): 5...70 °C; general: 0...70 °C (max. 24 °C under ambient).
- Humidity range: 10...80 % RH.
- Max. heat compensation at 40 °C: 400 W. Nominal power: 1.2 kW. Rated voltage: 200...230 V / 100...120 V; 50/60 Hz; fuse 16.0 A; single-phase.
- Interior volume: 1,020 L. Interior (W×H×D): 973 × 1,250 × 836 mm. Net housing (W×H×D): 1,254 × 1,925 × 1,146 mm.
- Net weight (empty): 322 kg. Standard shelves (std./max.): 2 / 15. Load per rack: 45 kg. Permitted total load: 150 kg.
- Temperature uniformity/variation examples: 25 °C / 60 % RH spatial uniformity 0.1 ± K; 40 °C / 75 % RH spatial uniformity 0.5 ± K.
- Recovery times (door opened 30 s): 25 °C/60 % RH = 5 min; 40 °C/75 % RH = 15 min.
- Energy consumption: 90 Wh/h (25 °C / 60 % RH); 135 Wh/h (40 °C / 75 % RH).
- Interfaces & options: Ethernet, internal data logger + USB export; many accessories and qualification/calibration services available (access ports, LED kits, PURE AQUA, APT-COM™ software, IQ/OQ/PQ packages).