Turbidity water quality sensor DBA-800A
blue-green algaeindustrial

Turbidity water quality sensor - DBA-800A - DABECO Co.,Ltd - blue-green algae / industrial
Turbidity water quality sensor - DBA-800A - DABECO Co.,Ltd - blue-green algae / industrial
Add to favorites
Compare this product

Characteristics

Measured parameter
turbidity, blue-green algae
Domain
industrial

Description

Product Overview
The Blue-green Algae sensor uses the characteristic that blue-green algae has absorption peak and emission peak in the spectrum to emit monochromatic light of specific wavelength into the water. Blue-green algae in the water absorbs the energy of monochromatic light, releasing monochromatic light of emission peak of another wavelength. The light intensity emitted by blue-green algae is proportional to the content of blue-green algae in water. This product is widely used in the on-line monitoring of blue-green algae imported from waterworks, drinking water source and aquaculture, as well as the on-line monitoring of blue-green algae in different water bodies such as surface water and landscape water.

Application
  • Blue-green algae online monitoring of water plant inlet, drinking water source, aquaculture, etc.
  • Blue-green algae online monitoring of different water bodies such as surface water, scenic water, etc.

Specification
Specifications
Details
Measurement Range
0-300,000 cells/mL (customizable)
Detection Limit
200 cells/mL
Measurement Accuracy
±10% of the signal level corresponding value of 1ppb Rhodamine B Dye
Repeatability
±3%
Resolution
20 cells/mL
Pressure Range
≤0.4Mpa
Requirements
Suggest a multipoint monitoring for the distribution of blue-green algae in water is very uneven. Water turbidity is below 50NTU.
Materials
Body: SUS316L + PVC; Cover: POM; Cable: PUR
Power Supply
9~36VDC
Communication Protocol
MODBUS RS485
Storage Temperature
-15~50℃
Operating Temperature
0~45℃ (not freeze)
Weight
0.55KG
Level of Protection
IP68/NEMA6P
Cable length
Standard: 10 m, the maximum can be extended 100m

Technical Characteristics / Specifications
  • Measurement Range: 0-300,000 cells/mL (customizable)
  • Detection Limit: 200 cells/mL
  • Measurement Accuracy: ±10% of the signal level corresponding value of 1ppb Rhodamine B Dye
  • Repeatability: ±3%
  • Resolution: 20 cells/mL
  • Pressure Range: ≤0.4Mpa
  • Materials: Body: SUS316L + PVC; Cover: POM; Cable: PUR
  • Power Supply: 9~36VDC
  • Communication Protocol: MODBUS RS485
  • Storage Temperature: -15~50℃
  • Operating Temperature: 0~45℃ (not freeze)
  • Weight: 0.55KG
  • Level of Protection: IP68/NEMA6P
  • Cable length: Standard 10 m, extendable up to 100m

Catalogs

No catalogs are available for this product.

See all of DABECO Co.,Ltd‘s catalogs

Other DABECO Co.,Ltd products

water quality sensors and controllers

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.