Product OverviewThis low-range turbidity sensor directs parallel light from its light source into the internal water sample, where the light is scattered by suspended particles. A silicon photocell receiver—immersed in the sample—captures the scattered light at a 90° angle to the incident beam, and turbidity values are calculated based on the correlation between this 90° scattered light and the incident light intensity. Widely applied for online turbidity monitoring, it is suitable for waterworks (filtration pre- and post-monitoring, product water), direct drinking water systems, as well as industrial scenarios including circulating cooling water, filtered effluent, and reclaimed water reuse systems.
Measurement PrincipleThis low-range turbidity sensor guides parallel light from its light source downward into the water sample. Suspended solids in the water scatter the light, and a silicon photocell receiver (immersed in the sample) captures the scattered light at a 90° angle to the incident beam. Turbidity values are calculated based on the correlation between the 90° scattered light and the incident light intensity.
Key Features- Stable measurement data with excellent repeatability.
- Simple cleaning and low maintenance requirements.
- Continuous-reading scattering turbidity meter, specialized for low-range turbidity monitoring.
- Plug-and-play design for quick, easy on-site installation.
Applications- Online turbidity monitoring for waterworks (filtration processes, product water) and direct drinking water systems.
- Online turbidity monitoring for recirculating cooling water, filter water in industrial production, and water reuse systems.
Product Specifications| Specifications | Details |
| Measurement Range | 0.001-100NTU |
| Measurement Accuracy | The deviation of reading in 0.001~40NTU is ±2% or ±0.015NTU (whichever is maximum); it is ±5% in the range of 40-100NTU. |
| Repeatability | ±2% |
| Resolution Ratio | 0.001~0.1NTU, depends on different ranges |
| Flow Rate | 300mL/min≤X≤700mL/min |
| Pipe Fitting | Injection Port: 1/4NPT; Discharge Outlet: 1/2NPT |
| Material | Machine: ABS+SUS316L; Sealing Element: Acrylonitrile Butadiene Rubber; Cable: PUR |
| Power supply | 13VDC |
| Communication protocol | MODBUS RS485 |
| Storage Temperature | -15~60℃ |
| Operating Temperature | 0~45℃ (not freeze) |
| Weight | 2.1KG |
| Level of Protection | IP65 (Indoor) |
| Cable Length | Standard: 3 m, the tungsten lamp version maximum can be extended 7 m |
Technical Characteristics / Specifications- Measurement Range: 0.001-100NTU
- Accuracy: ±2% or ±0.015NTU (0.001~40NTU), ±5% (40-100NTU)
- Repeatability: ±2%
- Resolution: 0.001~0.1NTU
- Flow Rate: 300-700mL/min
- Pipe Fitting: 1/4NPT (injection), 1/2NPT (discharge)
- Material: ABS+SUS316L, Acrylonitrile Butadiene Rubber, PUR cable
- Power Supply: 13VDC
- Communication: MODBUS RS485
- Storage Temp: -15~60℃
- Operating Temp: 0~45℃ (not freeze)
- Weight: 2.1KG
- Protection: IP65 (Indoor)
- Cable Length: 3m standard, up to 7m (tungsten lamp version)