Light scattering turbidity sensor DAW-TR20C-S
compacthigh-accuracyModbus RTU

Light scattering turbidity sensor - DAW-TR20C-S - DABECO Co.,Ltd - compact / high-accuracy / Modbus RTU
Light scattering turbidity sensor - DAW-TR20C-S - DABECO Co.,Ltd - compact / high-accuracy / Modbus RTU
Add to favorites
Compare this product

Characteristics

Technology
light scattering
Other characteristics
high-accuracy, Modbus RTU, RS-485, for water, compact, 4-20mA

Description

Product Overview
Online monitoring solution for drinking water quality. This pressurized turbidity sensor uses 90° laser nephelometry (660 nm) to detect scattered light from suspended particles and convert it to turbidity values via calibration. Key advantages for industrial drinking-water systems include a pressure-rated, sealed flow path that reduces bubbles and eases sample recovery, a compact footprint for straightforward integration, and a self-cleaning optical window to extend maintenance intervals.

Industry Applications
  • Plant outlet water
  • Secondary water supply
  • Distribution network endpoint water (tap water)
  • Direct drinking water
  • Membrane‑filtered water
  • Swimming pool water
  • Surface water

Product Features
  • Pressure-rated flow channel to reduce bubble formation
  • Sealed flow path to facilitate sample recovery
  • Compact size for easy system integration
  • Self-cleaning optical window to extend maintenance intervals
  • Ultra-low detection limit: 0.005 NTU
  • High accuracy in low-range measurements (±2% or 0.02 NTU, whichever is greater; Formazin, 25 °C)
  • Multiple selectable ranges: 0–1 / 20 (default) / 100 NTU; extended monitoring up to 2000 NTU
  • Outputs: RS485 Modbus-RTU; optional 4–20 mA
  • Ingress protection: IP68
  • Operating temperature: -5 to 45 °C
  • Sample flow: 100–1000 mL/min (recommended 500 mL/min)

Technical Specifications
Operating Voltage: DC 24 V (supports 9–28 V wide input)
Power: 5 W (peak)
Measurement Method: 90-degree laser nephelometry
Light Source: 660 nm laser
Range: 0–1 / 20 (default) / 100 NTU (multiple ranges; extended monitoring up to 2000 NTU)
Resolution: 0.0001 NTU (<1 NTU), 0.001 NTU (≥1 NTU)
Limit of Detection: 0.005 NTU
Accuracy: ±2% or 0.02 NTU (whichever greater) based on Formazin at 25 °C
Operating Temperature: -5 to 45 °C
Operating Mode: Pressurized flow-through
Operating Pressure: ≤ 0.6 MPa
Sample Flow: 100–1000 mL/min (recommended: 500 mL/min)
Dimensions (W × H × D): 192 × 147 × 98 mm (including fittings and backplate)
Enclosure Material: ABS
Outputs: RS485 Modbus-RTU, optional 4–20 mA
Note: Manufacturer reserves the right to modify specifications for product improvements.

Catalogs

No catalogs are available for this product.

See all of DABECO Co.,Ltd‘s catalogs

Other DABECO Co.,Ltd products

water quality sensors and controllers

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.