OverviewValProbe RT loggers with bendable sensors cover a wide temperature range from -85°C to +400°C. They are offered in single-, dual- and five-channel bendable-sensor versions. Precision platinum RTD sensing provides high measurement accuracy while the logger electronics are designed for extended battery life. Operator-programmable sample rates start at 1 second and each sensor can store up to 100,000 data points. Antenna diversity ensures robust RF communication reliability.
Available models / Variants- 1- and 2-Channel Bendable Logger: sensor temperature range -85°C to 400°C. Single and dual bendable sensors available. Sensor length options: 12", 24", 36". Tip diameter: 2.4 mm (0.095").
- 5-Channel Bendable Logger: sensor temperature range 0°C to 400°C. Five bendable sensors. Sensor length: 24". Tip diameter: 2 mm (0.078").
Applications- Steam sterilizers
- Dry heat sterilizers
- Steam-in-Place (SIP) processes
- Water cascade/fall sterilizers
- Incubators
- Stability chambers
- Freezers
- Freeze drying / lyophilization processes
- Pressure vessels and reactors
Technical summary (excerpt)- Sensor type: Single, Dual or Five precision RTD sensors
- Sensor lengths: 12", 24", 36" (305, 610, 915 mm) — 5-channel uses 24" (610 mm)
- Tip diameters: 0.095" (2.4 mm) and 0.078" (2 mm)
- Measurement range: -85°C to 400°C (some 5-channel variants 0°C to 400°C)
- Accuracy: -85°C to 140°C: ±0.1°C; 140°C to 400°C: ±0.25°C
- Logger dimensions: H 1.9" (48 mm); Ø 1.4" (36 mm)
- Materials: Stainless steel 316L and PEEK
- Battery: Field-replaceable 3.6 V lithium cell
- Sampling rate: Operator-programmable starting at 1 s
- Data storage: Up to 100,000 samples per sensor in non-volatile memory
- Real-time clock accuracy: < 15 sec/day
- Calibration: NVLAP / DAkkS calibration available; automated user verification capability
- Sensing element: Precision platinum RTD
- Environmental ranges: Operational temperature (device body) -85°C to 140°C; Pressure 0–5 bar absolute; Humidity 0–100% condensing
- Regulatory: UL, CE
Key features- Wide measurement range from -85°C to +400°C
- Available as single-, dual- and five-channel bendable-sensor loggers
- High-accuracy platinum RTD sensing
- Long service life with field-replaceable battery
- Programmable sample rates and large non-volatile data storage
- Robust RF communication using antenna diversity
- Corrosion-resistant stainless steel 316L and PEEK construction
- NVLAP / DAkkS calibration and automated user verification options
Characteristics / Technical specifications- Data type: Temperature
- Distinctive feature: Bendable sensors
- Temperature limits: Max. 400 °C (752 °F); Min. -85 °C (-121 °F)
- Sensor configurations: Single, Dual, Five (model-dependent)
- Sensor lengths: 12", 24", 36" (305, 610, 915 mm) — 5-channel: 24" (610 mm)
- Tip diameters: 0.095" (2.4 mm) and 0.078" (2 mm)
- Device measurement range: -85°C to 400°C (some 5-channel variants 0°C to 400°C)
- Accuracy: ±0.1°C (-85°C to 140°C); ±0.25°C (140°C to 400°C)
- Logger dimensions: H 1.9" / Ø 1.4" (48 mm / 36 mm)
- Materials: Stainless steel 316L, PEEK
- Battery: Field-replaceable 3.6 V lithium
- Sampling rate: From 1 s (programmable)
- Memory: 100,000 samples per sensor, non-volatile
- RTC accuracy: < 15 sec/day
- Calibration: NVLAP / DAkkS
- Sensing element: Precision platinum RTD
- Environmental: Device body temperature -85°C to 140°C; Pressure 0–5 bar abs.; Humidity 0–100% condensing
- Compliance: UL, CE