Autoclave validation thermocouple wire
Kapton® insulatedfor validationstranded

Autoclave validation thermocouple wire - Kaye - Kapton® insulated / for validation / stranded
Autoclave validation thermocouple wire - Kaye - Kapton® insulated / for validation / stranded
Autoclave validation thermocouple wire - Kaye - Kapton® insulated / for validation / stranded - image - 2
Autoclave validation thermocouple wire - Kaye - Kapton® insulated / for validation / stranded - image - 3
Add to favorites
Compare this product

Characteristics

Options and accessories
for autoclave validation, for validation, stranded, Kapton® insulated
Temperature

Max.: 350 °C
(662 °F)

Min.: -273 °C
(-459 °F)

Description

Description The correct thermocouple sensor for the target temperature range and application is a critical element of your validation system. Kaye offers a range of Type T insulated thermocouple wires including an Ultra-Premium Autobond construction that eliminates the outer Teflon jacket and molds the inner conductors to prevent moisture and steam leakage. Options include Ultra-Premium Autobond, Ultra-Premium with Kapton jacket and a cost-effective Premium grade.

  • Autobond construction removes outer Teflon jacket to prevent moisture/steam ingress.
  • Ultra-Premium grade produced for maximum purity and uniformity.
  • Premium grade available for less critical applications at a lower cost.
  • Quality control: every spool tested and shipped with a Certificate of Conformance to traceable specifications.
  • Insulated Type T wires offered in 3-strand or 7-strand constructions; supplied uncut on spools or cut to length with Teflon tips.
  • Standard measurement accuracy: ±0.25°C (typical); Ultra-Premium accuracy: ±0.1°C.
  • Ultra-Premium Kapton option for applications up to 350°C.


Applications

  • Steam sterilizers (autoclaves)
  • Dry heat sterilizers
  • Steam-in-place (SIP) systems
  • Water cascade/fall sterilizers
  • Incubators (including neonatal incubators)
  • Stability chambers
  • Freezers
  • Freeze dryers / lyophilization
  • Process and laboratory vessels


Technical specifications

  • Product family: Kaye Thermocouples
  • Sensor type: Thermocouple, Type T
  • Wire grades: Ultra-Premium Autobond; Ultra-Premium Kapton; Premium grade also available
  • Construction: Insulated Type T, 3-strand or 7-strand
  • Availability: uncut spools or cut-to-length with Teflon tips
  • Example models: 3-Strand K0258; 7-Strand K0255; 7-Strand (AL) K0259; 7-Strand (AS) K0263; 7-Strand K0250
  • Gauge options: 27 AWG, 26 AWG, 22 AWG
  • Measurement accuracy (general): ±0.25°C; Ultra-Premium: ±0.1°C
  • Manufacturer examples: @121°C ±0.3°C; @300°C ±1.2°C
  • Approximate diameter range: 0.042” (1.07 mm) to 0.120” (3.03 mm)
  • Maximum group variation: @40°C ±0.05°C max; @121°C ±0.1°C max; @300°C ±0.3°C max
  • Ultra-Premium Kapton jacket rated to 350°C