The Oxygen Index (OI) gives us an accurate way of measuring the minimum concentration of oxygen percentage, that will just support the material flaming under a constant airflow and ambient temperature.
MOTIS’s latest Auto Oxygen Index Tester helps the user to perform a oxygen index test in a quicker, more accurate, and easier method.
• Auto flow control.
• Auto concentration adjustment.
The Auto Oxygen Index consists of:
• A compact integrated body tester.
• A heat-resistant glass tube, dimension of φ 75 by 500mm in height.
• The glass tube is clamped by mechanical arm for motorized lifting.
• Oxygen concentration transducer, with the following characteristics:
a) Measurement range of 0 to 100%
b) Linearity less than 0.1%
c) Repeatability less than 0.1%
d) Intrinsic error less than 0.1%
e) Zero stability less than 0.1% O2 per week
f) Response time T90 less than 3 secs at the flow of 80ml/min.
• Digital display of oxygen percentage, resolution at 0.1%
• Mass flow meters for oxygen and nitrogen flow control.
• Digital display of oxygen and nitrogen flow, resolution at 0.1 L/min.
• Auto zero & span calibration.
• Auto gas concentration & flow adjustment.
• Stainless steel made specimen holders.
• Portable flame igniter, easy to use.