OverviewThe Fluke 5560A is a multi-product laboratory calibrator designed to cover broad electrical workloads with high accuracy. It supports DC/AC voltage and current, resistance, capacitance, inductance and temperature, with options for oscilloscope calibration and high-current expansion.
Key features- Large color touch-screen with Visual Connection Management for guided connections and error reduction
- Calibrates AC/DC voltage and current, resistance, capacitance, inductance and temperature sensors (thermocouples and RTDs)
- Oscilloscope output option up to 2 GHz (amplitude, edge, bandwidth, time markers, impedance, overload)
- 30 A native continuous current output; expandable to 360 A direct or up to 6000 A with Fluke coils and 52120A amplifier
- Fast settling times, isolated terminal block and multi-unit synchronization modes
What’s in the box- ISO/IEC 17025 accredited calibration certificate with data (traceable)
- Basic test leads and adapters
- Rugged transport case
Product family and model differentiation- 5560A: Next-generation high-performance calibrator — widest workload coverage (bench 6.5-digit DMMs, oscilloscope up to 2 GHz, synthesized inductance, extended temperature connector support)
- 5550A: Performance model optimized for benchtop DMMs and oscilloscopes up to 1 GHz; improved TURs and 30 A continuous output
- 5540A: Standard model for common handheld and bench workloads (oscilloscopes to 600 MHz), optimized for cost of ownership
- 5530A: Basic calibrator for handheld DMMs and power devices (no oscilloscope capability)
Target workload & comparative capabilitiesTarget Workload | 5560A | 5550A | 5540A | 5530A
4.5-digit bench/handheld DMM (eg Fluke 87V) | TUR 4:1 | TUR 4:1 | TUR 4:1 | TUR 4:1
5.5-digit bench DMM (eg Fluke 8808A) | TUR 4:1 | TUR 4:1 | n/a | n/a
6.5-digit bench DMM (eg Keysight 34460A) | TUR 4:1 | n/a | n/a | n/a
Notes / ApplicationsScope bandwidth options, fast edge performance and time-marker accuracy support comprehensive oscilloscope calibration. Multi-unit sync, AC power/phase/PF and dual-output modes increase workload flexibility for lab and service environments.
Technical specifications (selection)- DC Voltage: 0 V to ±1020 V — example best performance 6 ppm (5560A)
- DC Current: 0 A to ±30.2 A continuous — example 78 ppm (5560A)
- AC Voltage: 1 mV to 1020 V; 3 Hz to 500 kHz — example 110 ppm (5560A)
- AC Current: 10 µA to 30.2 A continuous; 3 Hz to 30 kHz — example 120 ppm (5560A)
- Resistance: 0 Ω to 1.2 MΩ — example 19 ppm (5560A)
- Capacitance: 0 pF to 120 mF — example 0.09% (5560A)
- Inductance: 13 µH to 120 H — example 0.09% (5560A)
- Temperature: 17 thermocouples and 10 RTDs supported; ITS-90/IPTS-68 — example 0.09 °C TC, 0.002 °C RTD
- Oscilloscope output: leveled sine bandwidth option up to 2 GHz; vertical amplitude 0 V to 120 V; fast edge <175 ps; time marker accuracy 2.5 ppm