Product OverviewThe CM series clamp-on ultrasonic flow sensors / flow meters are compact, non-invasive devices with embedded electronics. They attach to flexible plastic tubing to measure liquid flow rate and detect air bubbles and large solid particulates. As the sensor does not contact the fluid, CM models are suitable for hygienic or contamination-sensitive environments such as medical devices, biopharma, food & beverage, industrial automation and laboratories.
ApplicationsBlood purification devices: heart-lung machines (HLM), ECMO (Extracorporeal Membrane Oxygenation), organ perfusion ,organ transportation devices, CRRT, circulation devices.
Dialysis devices: hemodialysis, peritoneal dialysis for real-time flow monitoring and early warning of blockages or air bubbles.
Blood pumps and contrast media injection pumps: accurate flow measurement for pump verification and imaging delivery control.
Measurable LiquidsWater, blood, saline, oil, paint, contrast media, liquids with few or no solid particles.
Normal Saline (0.9% NaCl)
Glucose Injection / Dextrose Solution
Lactated Ringer's Solution
Whole Blood
Red Blood Cell (RBC) Suspension
Platelet Concentrate
Fresh Frozen Plasma (FFP)
Peritoneal Dialysis Solution
Hemodialysis Solution
Contrast media
Povidone-Iodine Solution
Ethanol / Isopropyl Alcohol
Heparin Sodium Solution
Sterile Water for Injection
Mannitol Injection
Medical Lubricant / Ultrasound Gel
Insulin Injection
Irrigation Solution
Preservative / Fixative Solution
Artificial TearsNormal Saline (0.9% NaCl)
Glucose Injection / Dextrose Solution
Lactated Ringer's Solution
Whole Blood
Red Blood Cell (RBC) Suspension
Platelet Concentrate
Fresh Frozen Plasma (FFP)
Peritoneal Dialysis Solution
Hemodialysis Solution
Contrast Agent / Radiopaque Contrast Medium
Povidone-Iodine Solution
Ethanol / Isopropyl Alcohol
Heparin Sodium Solution
Sterile Water for Injection
Mannitol Injection
Medical Lubricant / Ultrasound Gel
Insulin Injection
Irrigation Solution
Preservative / Fixative Solution
Product FeaturesProduct FeaturesCM sensors provide precise flow measurement across each model range, enabling process control and quality-sensitive applications where accurate volumetric data is required.
Non‑contact measurement reduces contamination risk, making CM sensors appropriate for medical, biopharma and food processing lines where fluid integrity is critical.
In addition to flow and volume measurement, CM sensors detect air bubbles and can be used for dosing verification and as a flow-switch in system automation.
- Compact Integrated Design
Embedded circuitry eliminates the need for external wiring and simplifies installation on production or test lines where space and fast deployment matter.
No BlockagesExternal clamping avoids in‑line obstructions and dead zones, ensuring continuous measurement without interrupting the fluid path.
Sensor CalibrationCalibration can be tailored to tubing material, dimensions, temperature and liquid type to optimize accuracy for specific installations.
Product SelectionPart # | Flow Range | Accuracy | Tubing OD | Tubing ID
CM-016 | Default: 1000 mL/min; Max: 2000 mL/min | 0–200 mL/min: ±6 mL/min; 200–2000 mL/min: ±3% | 5/32" (4 mm) | 1/8" (3.2 mm) or 4 mm
CM-019 | Default: 2000 mL/min; Max: 4000 mL/min | 0–200 mL/min: ±6 mL/min; 200–4000 mL/min: ±3% | 3/16" (4.8 mm) | 1/8" (3.2 mm) or 5 mm
CM-025 | Default: 3000 mL/min; Max: 6000 mL/min | 0–300 mL/min: ±9 mL/min; 300–6000 mL/min: ±3% | 1/4" (6.4 mm) | 1/8" (3.2 mm) or 6 mm
CM-038 | Default: 4000 mL/min; Max: 8000 mL/min | 0–500 mL/min: ±15 mL/min; 500–8000 mL/min: ±3% | 3/8" (9.5 mm) | 1/4" (6.4 mm) or 9 mm
CM-050 | Default: 5000 mL/min; Max: 10000 mL/min | 0–700 mL/min: ±21 mL/min; 700–10000 mL/min: ±3% | 1/2" (12.7 mm) | 3/8" (9.5 mm) or 13 mm
CM-063 | Default: 7000 mL/min; Max: 14000 mL/min | 0–900 mL/min: ±27 mL/min; 900–14000 mL/min: ±3% | 5/8" (15.9 mm) | 3/8" (9.5 mm) or 16 mm
CM-075 | Default: 10000 mL/min; Max: 20000 mL/min | 0–1200 mL/min: ±36 mL/min; 1200–20000 mL/min: ±3% | 3/4" (19.1 mm) | 1/2" (12.7 mm) or 19 mm
CM-100 | Default: 25000 mL/min; Max: 50000 mL/min | 0–2000 mL/min: ±60 mL/min; 2000–50000 mL/min: ±3% | 1" (25.4 mm) | 3/4" (19.1 mm) or 25 mm
Technical Specifications