Electrochemical chlorine sensor CC1
for water treatmentfor process measurementIP65

Electrochemical chlorine sensor - CC1 - DABECO Co.,Ltd - for water treatment / for process measurement / IP65
Electrochemical chlorine sensor - CC1 - DABECO Co.,Ltd - for water treatment / for process measurement / IP65
Add to favorites
Compare this product

Characteristics

Technology
electrochemical
Applications
for water treatment, for process measurement
Protection class
IP65
Other characteristics
for real-time measuring, with temperature sensor, maintenance-free
Measuring range (ppm)

Min.: 0 unit

Max.: 10 unit

Description

Product information
German-made CC1 free chlorine sensor adopts a membrane-covered amperometric potentiostatic 3-electrode system with built-in electronics. It measures free chlorine, features automatic temperature compensation, requires no zero-point adjustment, and has highly reduced pH dependence. It is mainly applied to swimming pool water, drinking water, and sea water.

Technical characteristics
This residual chlorine sensor offers good measurement stability, practical operation and maintenance, strong environmental adaptability, flexible signal and connection options, and complies with EMC and RoHS standards.

Special features
  • reduced dependence on pH
  • tolerates surfactants
  • requires little maintenance
  • tolerates low conductivity

Applications
  • Swimming-pool water
  • Drinking water
  • Service water
  • Process water
  • Sea water

Ambient conditions
  • Max. temperature: 45 °C
  • Max. pressure: 3 bar
  • Min. conductivity: 10 μS/cm

Technical Specification
Measurement Parameters: Free chlorine (based on isocyanuric acid with reduced dependence on pH-value)
Measuring Ranges/Resolution: 0.05…20.00 ppm: 0.01 ppm; 0.005…2.000 ppm: 0.001 ppm; 0.05…5.00 ppm: 0.01 ppm; 0.05…10.00 ppm: 0.01 ppm
Response time: T90 ≈ 2 minutes
Measuring water temperature: 0 ... +45 °C (no ice crystals in the measuring water)
Ambient temperature: 0 ... +55 °C
Temperature compensation: Automatically, by an integrated temperature sensor. Sudden temperature changes must be avoided.
Max. allowed working pressure: Operation with retaining ring in FLC: -0.5 bar (no pressure impulses and/or vibrations). Operation with retaining ring in FLC: -3 bar (no pressure surges and/or vibrations).
Flow rate (face velocity): approx. 15-30 l/h (33–66 cm/s) in flow cell; small flow rate dependence is given
pH range: pH 4 – pH 12 (highly reduced dependence on pH-value)
Run-in time: First start-up approx. 2 h
Zero point adjustment: Not necessary
Protection class: 5-pole M12 plug-on flange: IP68; 2-pole terminal with mA-hood: IP65
Material: Microporous hydrophilic membrane, PVC-U, PEEK, stainless steel 1.4571

Catalogs

W771-DCC1
W771-DCC1
2 Pages
W761-DCC1
W761-DCC1
2 Pages

Other DABECO Co.,Ltd products

Disinfectant sensor

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.