Flow rate

Max.: 0.03 m³/h
(1.059 ft³/h)

Min.: 0.015 m³/h
(0.53 ft³/h)

Description

Overview
The German-made WP10.1 hydrogen peroxide sensor uses a membrane-covered amperometric 2-electrode system to measure hydrogen peroxide (H₂O₂) across multiple ranges up to 200000 ppm. Designed for industrial water treatment including seawater, it operates in a pH range of 2–11 and offers mechanical durability and high resistance to surfactants and chemicals. The sensor includes automatic temperature compensation and is available with multiple flexible output options. With IP65/IP68 protection and calibration via on-device analytical determination, no zero-point adjustment is required.

Special features
  • Membrane-covered amperometric 2-electrode system
  • Mechanically robust membrane design
  • High tolerance to surfactants and aggressive chemicals
  • Automatic temperature compensation (integrated temperature sensor)
  • Multiple output versions and connector options (IP65/IP68)


Applications
  • Industrial water treatment plants, including seawater applications
  • Water disinfection processes using high H2O2 concentrations
  • Continuous process monitoring where wide measuring ranges are required


Process conditions
  • Max. measuring water temperature: 45 °C (no ice crystals in the measuring water)
  • Ambient temperature: 0 to +55 °C
  • Max. allowed working pressure (with retaining ring in flow cell): approx. -0.5 bar; specific mounting up to -1.0 bar under controlled conditions (avoid pressure surges and vibrations)
  • Flow rate (face velocity) in flow cell: approx. 15–30 L/h (small flow rate dependence)
  • pH range: pH 2 – pH 11
  • First start-up / run-in time: approx. 60 min
  • Zero point adjustment: Not necessary


Technical specification
  • Model: WP10.1
  • Measurement parameter: Hydrogen peroxide (H₂O₂)
  • Measuring ranges / resolution: 0.5–200 ppm: 0.1 ppm; 5–2000 ppm: 1 ppm; 0.005–2 % (20000 ppm): 0.001 % (10 ppm); 0.05–20 % (200000 ppm): 0.01 % (100 ppm)
  • Response time: T90 ≈ 12 min (sensor); temperature sensor t90 ≈ 8 min
  • Measuring water temperature: 0…+45 °C; Ambient temperature: 0…+55 °C
  • Temperature compensation: Automatic via integrated temperature sensor (max. change recommended 5 °C per hour)
  • Protection class: 5-pole M12 plug-on flange: IP68; 2-pole terminal with mA-hood: IP65
  • Materials: Elastomer membrane, PVC-U, stainless steel 1.4571
  • Calibration: Via on-device analytical determination
  • Run-in time: Approx. 60 min at first start-up