Formaldehyde analyzer GA315
exhaust gaslandfillfor monitoring gas emissions

Formaldehyde analyzer - GA315 - Fresenius Umwelttechnik GmbH  - exhaust gas / landfill / for monitoring gas emissions
Formaldehyde analyzer - GA315 - Fresenius Umwelttechnik GmbH  - exhaust gas / landfill / for monitoring gas emissions
Add to favorites
Compare this product

Characteristics

Measured entity
exhaust gas, formaldehyde
Application domain
for monitoring gas emissions, for regulatory, for power plants, landfill, for anaerobic digestion
Measured value
concentration
Configuration
for integration, online, rack-mount
Mode of use
automatic
Technology
NDIR
Other characteristics
continuous, real-time, sampling, with touchscreen, high-temperature, extractive, for 19'' racks

Description

The GA315 is a formaldehyde analyzer for hot online measurement of formaldehyde (CH2O) in exhaust gas at gas temperatures up to 300 °C, typically from gas engines running on biogas or landfill gas. The sample stays hot from the heated sample line to the measuring cell, so water-soluble formaldehyde is measured without the losses a sample gas cooler would cause.
Typical applications:
  • Formaldehyde emissions from gas engines at biogas and landfill gas plants
  • Formaldehyde emission testing on stationary engines, e.g. for limits under RICE NESHAP in the US
  • Readjusting engine settings directly during the measurement
  • Power generation and waste management
  • Engine manufacturers: exhaust gas development and testing
What operators get:
  • Results on site, without sending gas samples to a laboratory
  • Compensation for pressure, humidity and methane, so the methane content of engine exhaust does not distort the formaldehyde reading
  • NDIR method with vacuum reference technology and a cuvette heated to 85 °C
  • 7-inch color touch display with integrated PC
  • RS232 as standard; Profibus DP, Modbus TCP, EtherNet/IP, Profinet and up to 8 analog 4–20 mA outputs as options
Technical data (excerpt):
  • Gas component: formaldehyde (CH2O)
  • Measuring range: 0–200 mg/m³ or 0–160 ppm
  • Measurement uncertainty: < 5 % of reading
  • Detection limit: 4 mg/m³ or 3 ppm
  • Measurement interval: 20 s
  • Sample gas: up to 300 °C; heated sample inlet and heated sample line
  • Readiness: 5 min, full specification after 30 min
  • Housing: 19-inch rack, 400 mm deep, IP20, 17 kg
Full specifications: data sheet (as of 20.05.2025) under Catalogs.

VIDEO

Exhibitions

Meet this supplier at the following exhibition(s):

ACHEMA 2027
ACHEMA 2027

14-18 Jun 2027 Frankfurt am Main (Germany)

  • More information
    *Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.
    About DirectIndustry
    DirectIndustry is a marketplace that offers B2B players a complete experience, from product sourcing to purchase. DirectIndustry is an entity of VirtualExpo Group.