OverviewThis MIC100OZ ozone gas analyzer features an automatic zero calibration process (company invention patent). It is designed for continuous 24-hour online monitoring of ozone concentration at the outlet of ozone generators or in exhaust gas (a dehumidifier may be required).
Detection principleDual-path ultraviolet (UV) absorption method using an imported UV LED light source; supports continuous 24/7 online operation.
Measurement ranges (options)- 0-300 g/Nm3
- 0-200 g/Nm3
- 0-100 g/Nm3
- 0-50 g/Nm3
- 0-1000 ppm
Product featuresIP65-rated waterproof and dustproof housing for outdoor installation with waterproof power and signal aviation connectors. Built-in automatic zero calibration device uses reference light data and an internal PID controller to maintain zero stability; scheduled or gas-triggered zero calibration is available. Continuous 24-hour measurement capability for ozone generator outlets and exhaust streams.
Signal terminal outputSignal terminal outputs and related close-up images are provided in the product documentation and image section.
Technical specifications- Detection principle: Dual-path UV absorption, UV LED light source
- Measurement range: standard 0-300 g/Nm3; optional 0-200, 0-100, 0-50 g/Nm3; optional 0-1000 ppm
- Light source lifetime: approx. 50,000 hours (≈4–5 years)
- Sampling method: positive pressure sampling; negative pressure sampling optional
- Display: 7-inch touch screen
- Displayed parameters: ozone concentration, temperature, pressure
- Auxiliary functions: temperature compensation, pressure compensation
- Display units: g/Nm3, mg/NL, %WTAir, %WTO2 (optional)
- Resolution: 0.01 g/Nm3
- Gas flow rate: 0.3–0.7 L/min
- Input pressure: <0.1 MPa
- Concentration error: ≤ 0.5% FS
- Linearity error: ≤ 0.5% FS
- Zero drift: <±0.3% FS
- Response time: signal 0.3 s, display 0.3 s
- Operating temperature: -20 to 50 °C
- Interface: quick twist (stainless steel)
- Sampling aperture: Φ6 (6 mm × 4 mm)
- Communication: RS-485 Modbus RTU
- Output: 4-20 mA (active)
- Relay: high-position relay (NO COM NC), low-position relay (NO COM NC)
- Power supply: AC 110-220 V
- Dimensions (H×W×D): 350 × 270 × 150 mm
- Conversion note: 1 mg/L = 1 g/m3 = 467 PPM
- Warranty: 24 months (main sensor)
- Standard configuration (built-in): flowmeter, ozone exhaust catalyst, intake filter