Product overviewModel: OZ-150W
Detection principle: UV dual-path absorption method for analysis and detection of dissolved ozone concentration in ultrapure water and industrial process water (e.g. semiconductor industry).
Key features- Patented process with automatic UV light source adjustment to maintain real-time light stability.
- Dual-path UV absorption measurement principle for improved accuracy.
- Long-life UV LED light source (approx. 30,000–50,000 hours) designed for continuous operation.
- 4.3-inch touchscreen interface with selectable units (g/m3, mg/L, ppm).
Technical parameters (summary)Measuring ranges: 0–50 ppm; 0–100 ppm; 0–200 ppm; 0–300 ppm
Light source life: ~30,000–50,000 hours
Sampling methods: positive pressure sampling, negative pressure sampling (requires pumps, valves and flow meters at outlet)
Ozone contact materials: quartz, PTFE, PFA, FFPM (HF version: sapphire, PTFE, PFA, FFPM)
Display: 4.3-inch touchscreen; units: g/m3, mg/L, ppm (optional); resolution: 0.01 ppm
Sampling flow: 200–300 mL/min
Input pressure: <0.1 MPa
Concentration error: max 0.3% FS; linear error: max 0.2% FS; zero drift: <±0.3% FS
Operating temperature: 0–50°C
Interface: quick-twist (PTFE); sampling diameter: Φ6 (6mm×4mm)
Communication/output: RS-485; 4–20 mA (active)
Power supply: 220V AC (optional 24V DC)
Dimensions: 180 × 260 × 90 mm (H × W × D); Weight: 3 kg
Warranty: 24 months (host)
Process & diagramsProcess flow and product connection diagrams are provided to support system integration and installation.
Characteristics / specifications- Model: OZ-150W
- Detection principle: UV dual-path absorption
- Available ranges: 0–50, 0–100, 0–200, 0–300 ppm
- Resolution: 0.01 ppm
- Light source life: ≈30,000–50,000 h
- Sampling: positive/negative pressure (requires pumps/valves/flow meters)
- Ozone wetted materials: quartz, PTFE, PFA, FFPM (HF: sapphire option)
- Display: 4.3" touchscreen
- Units: g/m3, mg/L, ppm
- Sampling flow: 200–300 mL/min; input pressure: <0.1 MPa
- Accuracy: concentration error ≤0.3% FS; linear error ≤0.2% FS; zero drift <±0.3% FS
- Operating temperature: 0–50°C
- Interface: quick-twist PTFE; sampling diameter Φ6 (6mm×4mm)
- Communication: RS-485; analog output: 4–20 mA (active)
- Power: 220V AC (24V DC optional)
- Dimensions/weight: 180×260×90 mm / 3 kg
- Warranty: 24 months (host)