Product OverviewThe MIC600-OZ rack-mounted ozone gas concentration analyzer employs a dual-path ultraviolet absorption system with an intelligent lamp management module. The UV lamp reaches operational state rapidly after power-up. The analyzer uses a separated light pool design that prevents leakage, withstands high pressure and high-flow sampling impacts, and facilitates cleaning and maintenance. It is designed for continuous monitoring of ozone in pipelines, containers, ozone disinfection return-air ducts and cabinets, and ozone aging test chambers. Typical industries include pharmaceutical, chemical, municipal water, sewage treatment and others for measuring ozone generator outlet concentration and generator output.
Key Functional DescriptionThe instrument provides a color touch-screen interface with real-time display of ozone concentration, pressure and temperature, and supports automatic or manual zeroing. A high-concentration range (g/Nm3) accepts flow input to calculate and display ozone generator output in real time. Two relay outputs permit independent high and low alarm settings. Core components include a long-life UV light source and high-transmittance quartz plate to ensure zero stability and minimize drift. The pressure-resistant, impact-resistant separated light pool maintains gas flow and pressure during zeroing, enabling zeroing without interrupting ozone intake for continuous 24/7 operation.
Application Areas- Ozone generator manufacturers
- Municipal water treatment
- Industrial wastewater treatment
- Fine chemical industry
- Food and drinking water treatment
- Space disinfection
- Swimming pool disinfection
- Flavour synthesis industry
- Other industries using ozone generators
Ranges / Sampling / Application (summary)- 0–300 g/Nm3 | Sampling: positive pressure with pressure and temperature compensation | Application: real-time detection of ozone generator outlet concentration
- 0–100 g/Nm3 | Sampling: positive pressure with pressure and temperature compensation | Application: real-time detection of ozone generator outlet concentration
- 0–50 g/Nm3 | Sampling: positive or negative pressure with pressure and temperature compensation | Application: detection at ozone generator outlet or exhaust gas
- 0–1000 ppm | Sampling: positive or negative pressure with pressure and temperature compensation | Application: real-time detection inside cabinets or chambers
- 0–100 ppm | Sampling: positive or negative pressure with pressure and temperature compensation | Application: real-time detection in a space or room
Product Features- Test method: dual-path ultraviolet absorption with long-life UV source and high accuracy
- Measurement principle: photometric absorption based on Lambert–Beer law
- Light source: long-life UV lamp (253.7 nm); life expectancy ~30,000–50,000 hours
- Installation: indoor rack-mounted instrument with power and signal terminals
- Light pool: separated light pool technology — no leakage, high pressure resistance, resistant to high sampling flows
- Intelligent compensation: built-in temperature and pressure compensation and automatic light-source compensation
- Operation modes: manual or automatic zeroing with configurable interval
- Display units: selectable g/Nm3, mg/NL, %WT, ppm, mg/Nm3
- Interface: high-definition color touch screen; high-concentration mode accepts inlet gas mass flow input and shows real-time generator output
- Outputs: 4–20 mA, RS-485 communication, two alarm relay outputs, electronic flow meter input
- Standard accessories: anti-corrosion flow meter, ozone exhaust destroyer, air intake filter
Product Structure CompositionThe analyzer is built in a 19-inch 3U rack chassis with a separated light pool and modular components including lamp management system, flow and pressure interfaces, touch display and standard peripheral fittings. Product diagrams are available in the technical documentation.
Technical Specifications- Model (example): UVOZ-3000
- Commercial model: MIC600-OZ
- Measurement ranges: typical 0–100 ppm (customizable up to 0–1000 ppm); other ranges: 0–500 ppm (0.01 ppm); 0–1000 ppm (0.01 ppm); 0–50 ppm (0.001 ppm); 0–10 ppm (0.001 ppm)
- Light source life: approx. 30,000–50,000 h
- Sampling method: positive pressure sampling; negative pressure optional
- Display: 4.3-inch touch screen
- Displayed parameters: ozone concentration, temperature, pressure
- Auxiliary functions: temperature and pressure compensation
- Display units/resolution: ppm or mg/Nm3; resolution 0.01 ppm
- Gas flow: 0.5 L ± 0.2 L/min
- Input pressure: <0.1 MPa
- Concentration error: ≤ 0.5% FS (continuous); max 0.1% FS (intermittent zero)
- Linear error: ≤ 0.5% FS (continuous); max 0.1% FS (intermittent zero)
- Zero drift: < ±0.3% FS
- Response time: signal 0.03 s; display refresh 1 s (continuous); display every 5 s (intermittent)
- Ambient temperature: -20 to 50 °C
- Pipeline interface: quick screw (stainless steel)
- Sampling tube diameter: serial connection Φ8 (8 mm × 6 mm) optional; bypass Φ6 (6 mm × 4 mm)
- Communication: RS-485
- Analog output: 4–20 mA (active)
- Relay outputs: high and low alarm relays
- Power supply: AC 110–220 V
- Dimensions: 133 mm (H) × 482 mm (W) × 300 mm (D), 19-inch 3U chassis
- Notes: unit conversion guidance provided in product documentation (e.g., 1 mg/L = 1 g/m3 = 467 ppm)
- Warranty: 24 months (host)
- Standard built-in accessories: anti-corrosion flow meter, ozone exhaust destroyer, air intake filter