Product DescriptionThe UVOZ-180W Dissolved Ozone Analyzer measures dissolved ozone concentration in tap water, purified/pure water and ultrapure water using a UV dual-path absorption method. The instrument includes an automatic light-source compensation system that adjusts UV LED intensity in real time to maintain zero-point stability. It is designed for continuous 24-hour industrial operation and provides standard 4–20 mA and RS-485 (Modbus) outputs for PLC integration.
Product OverviewDirect photometric measurement using ultraviolet spectrophotometry (Lambert–Beer principle) allows reliable detection in low-turbidity and ultrapure environments. An HF-compatible version is available for use with 20% hydrofluoric acid ozone solutions. Wetted materials include PTFE, PFA, FFPM, quartz or sapphire to ensure chemical compatibility with ozone-containing water.
Product Features- Test method: Dual-path UV absorption with long-life light source for high measurement accuracy.
- Measurement principle: Photometric absorption following Lambert–Beer law.
- Selectable measurement ranges: 0–20, 0–50, 0–100, 0–120, 0–180, 0–200 ppm (some configurations up to 0–300 ppm).
- Light source: Long-life UV LED (≈253.7 nm) with light-source compensation and warranty period.
- Automatic compensation: Automatic zero adjustment and automatic light-source compensation to maintain a stable baseline.
- Operation: High-definition color touch-screen display; manual zeroing supported (requires clean, ozone-free water).
- Installation: Wall-mount aluminum housing, splash-proof (IP65) with mounting brackets supplied.
- Sampling: Supports positive-pressure and negative-pressure sampling. Recommended flow 200–300 mL/min; install throttle valve or flow meter at outlet when flow control is required.
- Outputs: 4–20 mA (active) and RS-485 (Modbus) for external control and data logging.
Product Structure & ConnectionsWall-mount enclosure with hook-style mounting and four mounting brackets. Display: high-definition touch screen (operator interface). Power input: 24 V DC (+ / -). Signal terminals (summary): RS485+ / RS485- (Modbus), 4–20 mA+ / 4–20 mA-, ZEROING+ / ZEROING-. Gas/compressed-air interface provided to prevent condensation at low temperatures. Sampling connections support branch positive-pressure or pump negative-pressure schemes; when split sampling is used, install flow meter at outlet.
Notes & Use Recommendations- Keep inlet water and sensor surfaces clean; perform zero calibration regularly (recommended every 2 weeks) using clean water without ozone or color.
- Do not install a throttle valve at the inlet to avoid bubble formation; place throttle/flow meter at the outlet if flow control is required.
- Ensure sampling pump, valves and outlet flow meter are properly sized for 200–300 mL/min recommended flow.
Technical Specifications- Model: UVOZ-180W
- Measurement ranges (examples): 0–20, 0–50, 0–100, 0–120, 0–180, 0–200 ppm (optionally 0–300 ppm)
- Light source life: ~30,000–50,000 hours
- Sampling method: positive-pressure sampling or negative-pressure (pump) sampling (outlet flow meter required)
- Materials in contact with sample: quartz, PTFE, PFA, FFPM, sapphire
- Display: 4.3-inch high-definition color touch screen
- Display units: mg/L or ppm (selectable)
- Resolution: 0.01 ppm
- Recommended sampling flow: 200–300 mL/min
- Input pressure: < 0.1 MPa
- Concentration error: ≤ 0.3% FS
- Linearity error: ≤ 0.2% FS
- Zero drift: < ±0.3% FS
- Operating temperature: 0–50 °C
- Sample interface: ferrule (PTFE) with optional 1/4 PFA interface
- Sampling tubing size: Φ6 (6 mm × 4 mm)
- Communication: RS-485 (Modbus)
- Analog output: 4–20 mA (active)
- Power supply: 24 V DC
- Dimensions: 150 × 200 × 80 mm (H × W × D)
- Weight: ≈ 3 kg
- Warranty: 24 months (host)