Product overviewThe UV OZ-180W Dissolved Ozone Water Concentration Monitor is a UV photometer for direct measurement of ozone in ultrapure or constant-turbidity water. A chemical-resistant variant (UV OZ-180W-HF) is available for 20% hydrofluoric acid ozone solutions. Wetted parts: PTFE, FFPM, PFA and quartz or sapphire. The unit provides a water inlet and outlet, a high-definition touch screen and signal outputs 4–20 mA and RS-485.
Measurement principle and operationDetection is by UV dual-path absorption based on the Lambert–Beer law for accurate photometric measurement. The analyzer uses a true dual-beam optical system and a long-life UV LED light source (253.7 nm). An automatic light-source adjustment system continuously regulates LED brightness against the reference channel to stabilize the zero point. Designed for continuous 24-hour operation. Light source warranty: 3 years.
Installation & sampling notesEnclosure: splash-proof aluminum, protection level IP65; wall-mountable using rear hook and supplied brackets. Recommended sampling flow: 200–300 mL/min. If flow control is needed, install throttle valve or flow meter at the outlet (never at the inlet). Sampling may be positive- or negative-pressure but must include pumps, valves and flow meters at the outlet. Keep inlet and internal sensor clean. Zero calibration requires clear water without ozone or color; calibration recommended every 2 weeks.
Product features- Test method: UV dual-path absorption for high accuracy
- Automatic zero adjustment and automatic light-source compensation
- Separate light pool design with no leakage, high pressure resistance and resistance to flow-sampling impact
- High-definition color touch-screen for data display and operation
- Versions: standard for tap/purified water (UV OZ-180W) and chemical-resistant version for HF solution (UV OZ-180W-HF)
Technical specifications- Model: OZ-180W
- Measuring ranges (variants): 0–20 ppm; 0–50 ppm; 0–100 ppm; 0–120 ppm; 0–180 ppm; 0–200 ppm (alternative listings: 0–50 ppm; 0–100 ppm; 0–200 ppm; 0–300 ppm)
- Light source life: approx. 30,000–50,000 hours
- Sampling: positive-pressure or negative-pressure (requires pumps/valves/flow meters)
- Ozone contact materials: quartz or sapphire; PTFE; PFA; FFPM (chemical-resistant variant uses appropriate materials)
- Display: 4.3-inch high-definition touch screen
- Units: mg/L or ppm (optional)
- Resolution: 0.01 ppm
- Sampling flow rate: 200–300 mL/min
- Input pressure: < 0.1 MPa
- Concentration error: max 0.3% FS
- Linearity error: max 0.2% FS
- Zero drift: < ±0.3% FS
- Ambient temperature: 0–50 °C
- Interface: ferrule (PTFE) or optional 1/4 PFA
- Sampling tubing size: Φ6 (6 mm × 4 mm)
- Communication: RS-485
- Output: 4–20 mA (active)
- Power: 24 VDC (220 VAC optional)
- Protection: splash-proof aluminum enclosure, IP65
- Dimensions: 150 mm (H) × 200 mm (W) × 80 mm (D)
- Weight: approx. 3 kg
- Operation: continuous 24-hour capability
- Maintenance: keep inlet clean and perform zero calibration approx. every 2 weeks using clear water without ozone or color