General descriptionPTM600-OZ is a microprocessor-based UV 254 nm dual-path photometer for continuous measurement of ozone in air or oxygen. The analyzer monitors measurement and reference UV signals, cuvette temperature and pressure, and provides real-time ozone concentration on a 4.3-inch touchscreen with built-in temperature and pressure compensation. For high-concentration ranges (g/Nm3) it supports a flow input to display ozone generator output in real time. The instrument uses a long-life UV light source and high-transmittance quartz to minimize zero drift and enable continuous 24 h operation without closing the ozone inlet for zeroing. Display units can be switched between %wt/wt, g/Nm3, ppmv (AQ: g/m3 or ppm).
Additional modifiable parameters- Pressure display unit (kPa)
- Alarm thresholds (high/low)
- Carrier gas selection: air or oxygen (including PSA)
- RS-485 communication parameters
- 4–20 mA analog output settings
Installation and samplingUse 6 × 4 mm PTFE tubing for sample connection via the sample gas filter (IN). Measurement lag depends on sample flow rate, tubing length, tubing cross-section (recommended 6×4 mm) and instrument internal response. At 0.2–1 L/min and 2 m tubing, expected lag ≈ 0.4–2 s. Sampling method: positive pressure sampling.
Display and operation4.3-inch touchscreen displays ozone concentration, temperature and pressure. Auxiliary functions: temperature and pressure compensation. Selectable display units: g/Nm3, mg/NL, %wt/wt, ppmv. Display resolution: 0.01 g/Nm3.
Performance and environmental conditionsGas flow rate: 0.5 L ± 0.2 L/min. Input pressure: < 0.1 MPa. Concentration accuracy: max 0.5% FS. Linearity error: max 0.5% FS. Zero drift: < ±0.3% FS. Response time: ≤ 1 s. Operating temperature: -30 to 80 °C.
Mechanical and connection featuresPipeline interface: quick-tightening (stainless steel). Bypass sampling diameter: Φ6 (6 mm × 4 mm). Communication: RS-485 (MODBUS). Signal output: 4–20 mA (active). Power supply: 24 V DC. Size: 120 × 160 × 60 mm. Note: 1 mg/L = 1 g/m3 = 467 ppm. Warranty: 24 months (host). Standard configuration includes anti-corrosion flow meter, ozone exhaust destructor and intake dust filter.
Product features- Patented automatic zero-point adjustment and dual-path UV absorption for stable baseline.
- High-accuracy measurement using dual-path UV absorption (Lambert–Beer law).
- Long-life UV LED light source with lens; lifetime ≈ 30,000–50,000 h; 2-year warranty on light source.
- Recommended for indoor installation; equipped with waterproof aviation connectors.
- Separate sealed light pool design resistant to leaks, pressure and high-flow sampling impact.
- Intelligent compensation: built-in temperature and pressure compensation and display.
Power input and signal terminals(Power input)
(01) 24V DC+ | (02) 24V DC-
(Signal terminal output)
(01) RS485+ (MODBUS) | (03) 4-20MA+
(02) RS485- (MODBUS) | (04) 4-20MA-
Technical specifications- Model: PTM600-OZ
- Detection principle: UV dual-path absorption (UV 254 nm)
- Measurement ranges (options): 0–300 g/Nm3; 0–200 g/Nm3; 0–100 g/Nm3; 0–50 g/Nm3
- Display: 4.3-inch touchscreen
- Display units: g/Nm3, mg/NL, %wt/wt, ppmv
- Display resolution: 0.01 g/Nm3
- Flow rate: 0.5 L ± 0.2 L/min
- Input pressure: < 0.1 MPa
- Concentration accuracy: max 0.5% FS; linear error max 0.5% FS
- Zero drift: < ±0.3% FS
- Response time: ≤ 1 s
- Operating temperature: -30 to 80 °C
- Bypass sampling diameter: Φ6 (6 mm × 4 mm)
- Communication: RS-485 (MODBUS)
- Signal output: 4–20 mA (active)
- Power supply: 24 V DC
- Size: 120 × 160 × 60 mm
- Light source lifetime: approx. 30,000–50,000 hours
- Warranty: 24 months (host)
- Standard accessories: anti-corrosion flow meter, ozone exhaust destructor, intake dust filter