Product OverviewThe UVOZ-1100D is a desktop ozone gas concentration analyzer using a UV dual-path absorption detection principle for real-time measurement of ozone in enclosures or exhaust streams. Measurement ranges: 0-100 ppm (standard) and optional 0-1000 ppm. Designed for continuous operation with automatic zero calibration and negative-pressure pump sampling to minimize zero drift and ensure stable readings.
Key FeaturesTouchscreen operation with real-time display of ozone concentration, pressure and temperature compensation and display. Core components include a long-life UV light source and high-transmittance quartz plate to maintain zero stability. Independent pressure-resistant light pool design preserves gas flow and pressure during zeroing, allowing zero calibration without interrupting ozone intake for 24/7 operation.
Application AreasSuitable for ozone generator manufacturers, municipal water treatment, industrial wastewater treatment, fine chemicals, food and drinking water industry, space disinfection, swimming pool disinfection, flavor synthesis and other sectors using ozone generators.
Range / Sampling / Application (table)range | Sampling method | Application areas
0-1000ppm | Positive or negative pressure sampling, with pressure and temperature compensation | Real-time detection of ozone concentration in enclosures/exhaust.
0-100ppm | Positive or negative pressure sampling, with pressure and temperature compensation | Real-time detection of ozone concentration in spaces or rooms.
Product Features (details)Test method: UV absorption; long-life light source; high measurement accuracy.
Measurement principle: Based on Lambert-Beer law using photometric absorption.
Light source system: Long-life UV light source system; warranty specified in product details.
Installation: Indoor mounting; equipped with power and signal terminals.
Light pool system: Separate light pool technology, leak-free, high pressure resistance and resistant to high-flow sampling impacts.
Intelligent compensation: Built-in temperature and pressure compensation and display, with automatic light source compensation.
Operation: Manual and automatic zeroing options; automatic zeroing interval configurable.
Display units: ppm, mg/Nm3 optional.
Data display: High-definition color touchscreen.
Output: 4-20mA, RS485 communication, relay alarm outputs.
Standard accessories: Ozone exhaust destructor, air intake filter.
Technical specifications- Model: UVOZ-1100D
- Measurement range: 0-100ppm (standard)
- Optional range: 0-1000ppm
- Light source life: approx. 20,000–30,000 hours
- Sampling method: negative pressure sampling (positive or negative pressure sampling available)
- Display interface: 4.3-inch touchscreen
- Display unit: ppm, mg/Nm3 (optional)
- Display resolution: 0.01 ppm
- Sampling gas flow: 200–800 mL/min (adjustable)
- Input pressure: atmospheric pressure 101 kPa
- Concentration error: Max 0.3% FS (intermittent zero sampling)
- Linearity error: Max 0.3% FS (intermittent calibration zero sampling)
- Zero drift: <±0.3% FS
- Response time: signal 0.03 s; display update every 5 s (intermittent calibration zero sampling)
- Ambient temperature: -20 to 50 °C
- Piping interface: Φ6 quick-screw connector (with air inlet filter)
- Bypass sampling calibre: Φ6 (6mm*4mm)
- Communication mode: RS-485
- Output mode: 4-20mA (active)
- Relay signal: high alarm relay, low alarm relay
- Power supply: 24V DC
- Dimensions: 160 mm (W) x 90 mm (H) x 330 mm (D)
- Notes: 1 mg/L = 1 g/m3 = 467 PPM
- Free warranty: 12 months (main unit)
- Standard accessories: built-in ozone exhaust destructor, standard air intake filter