Ozone analyzer UVOZ-1100
for ambient airindustrialconcentration

Ozone analyzer - UVOZ-1100 - Beijing Zetron Technology Co., Ltd. - for ambient air / industrial / concentration
Ozone analyzer - UVOZ-1100 - Beijing Zetron Technology Co., Ltd. - for ambient air / industrial / concentration
Ozone analyzer - UVOZ-1100 - Beijing Zetron Technology Co., Ltd. - for ambient air / industrial / concentration - image - 2
Add to favorites
Compare this product

Characteristics

Measured entity
ozone
Application domain
for ambient air, industrial
Measured value
concentration, temperature, pressure
Configuration
wall-mounted
Mode of use
automatic, manual
Technology
UV-DOAS
Other characteristics
continuous, real-time, RS485, with touchscreen

Description

Product overview
The UVOZ-1100 wall-mounted ozone gas concentration analyzer provides continuous online measurement of ozone in industrial environments, pipelines, containers and enclosed spaces. It combines UV dual-path absorption with separated light pool technology and an intelligent lamp management system for fast readiness, stable baseline and simplified maintenance.

Detection principle & operation
Detection principle: UV dual-path absorption (Lambert-Beer photometric absorption).
Operation interface: 4.3-inch color touch screen displaying ozone concentration, temperature and pressure in real time. Supports manual and configurable automatic zeroing times.

Key product features
  • UV dual-path absorption with long-life UV light source for reliable photometric measurement
  • Separated light pool design: leak-proof and resistant to pressure and high-flow sampling impact
  • Intelligent lamp management for rapid warm-up and stable zero point
  • Built-in temperature and pressure compensation; automatic light source compensation
  • Continuous 24-hour operation; zeroing can be performed without interrupting sampling
  • Selectable display units: ppm or mg/Nm3
  • Standard outputs: active 4–20 mA, RS-485, two relay alarm outputs
  • Standard accessories: ozone exhaust destructor and air intake filter

Applications
Designed for monitoring ozone generator outlets and ambient or pipeline ozone concentrations. Typical sectors: ozone generator manufacturers, municipal water treatment, industrial wastewater treatment, fine chemicals, food and beverage water treatment, space disinfection, swimming pool disinfection and flavor synthesis.

Range / Sampling / Application (summary table)
  • 0–1000 ppm — Positive or negative pressure sampling with pressure & temperature compensation — Real-time detection of ozone concentration in enclosures/boxes
  • 0–100 ppm — Positive or negative pressure sampling with pressure & temperature compensation — Real-time detection of ozone concentration in rooms or spaces

Model & parameter table (UVOZ-1100)
  • Model: UVOZ-1100
  • Measurement range: 0–100 ppm (standard); optional 0–1000 ppm
  • Light source life: approx. 20,000–30,000 hours
  • Sampling method: negative pressure sampling (positive/negative sampling with pressure & temperature compensation available)
  • Display interface: 4.3-inch touch screen
  • Displayed parameters: ozone concentration, temperature, pressure
  • Auxiliary functions: temperature compensation, pressure compensation
  • Display unit: ppm, mg/Nm3 (optional)
  • Display resolution: 0.01 ppm
  • Sampling gas flow: 500 mL/min
  • Input pressure: atmospheric (101 kPa)
  • Concentration error: max 0.3% FS (intermittent zero sampling)
  • Linearity error: max 0.3% FS (intermittent calibration zero sampling)
  • Zero drift: < ±0.3% FS
  • Response time: signal 0.03 s, display update every 5 s
  • Ambient temperature: -20 to 50 °C
  • Piping interface: pagoda connector (luer inlet filter)
  • Bypass sampling calibre: Φ6 (6 mm × 4 mm)
  • Communication: RS-485
  • Output: 4–20 mA (active)
  • Relay signals: high and low alarm relays
  • Power supply: 24 V DC
  • Dimensions: 160 × 330 × 110 mm (H × W × D)

Warranty & notes
Free warranty: 12 months for main unit. Standard built-in items: ozone exhaust catalyst and air intake filter. Follow recommended maintenance and calibration intervals for reliable operation.

Technical specifications
  • Detection method: UV dual-path absorption
  • Core components: long-life UV light source and high-transmittance quartz plates
  • Zeroing: automatic and manual; zeroing without stopping ozone intake
  • Measurement accuracy: concentration and linearity error up to 0.3% FS
  • Display resolution: 0.01 ppm
  • Sampling flow rate: 500 mL/min
  • Operating temperature: -20 to 50 °C
  • Power: 24 V DC
  • Outputs: 4–20 mA, RS-485, two relay alarm outputs
  • Piping interface: pagoda connector / luer inlet filter; bypass calibre Φ6
  • Dimensions: 160 × 330 × 110 mm (H × W × D)
  • Light source life expectancy: ~20,000–30,000 hours