Product nameDissolved ozone sensor (Model: OZ-300W UV; OZ-300W-HF available for 20% HF ozone solutions)
OverviewThe OZ-300W UV dissolved ozone analyzer is a UV photometer for direct measurement of dissolved ozone concentration in tap water, purified water, ultrapure water and constant-turbidity water. Wetted materials include PTFE, FFPM, PFA and quartz or sapphire depending on version. The instrument supports continuous 24-hour operation, features a true dual-beam optical system and a long-life UV LED light source (253.7 nm). It can be installed inline (parallel or series) on the dissolved-ozone pipeline and provides a 4.3-inch color touch screen with standard industrial outputs.
Key functional features- Detection principle: UV dual-path (dual-beam) absorption based on Lambert-Beer law.
- Automatic light-source adjustment: real-time LED/UV brightness adjustment using reference zero-point data to ensure zero stability.
- Long-life UV light source (253.7 nm) with service life ~30,000-50,000 h and warranty.
- Continuous 24-hour operation with enhanced stability due to dual-beam optical design.
- Wetted materials: PTFE, PFA, FFPM, quartz or sapphire (version-dependent).
- Output and communication: 4-20 mA analog output and RS-485 (Modbus) for PLC integration.
- User interface: 4.3-inch high-definition color touch screen.
- Enclosure: splash-proof aluminum, IP65; multiple mounting options (wall-mount with hook and brackets provided).
- Versions: standard OZ-300W UV for tap/purified water; OZ-300W-HF for 20% HF ozone solutions.
Installation & maintenance notesThe analyzer requires a sample inlet and outlet and a sampling flow rate of approximately 200-300 mL/min. If flow control is needed, install a throttle valve or flow meter at the outlet (never at the inlet) to avoid bubble generation; do not allow ozone water containing bubbles into the analyzer. Maintain cleanliness of the inlet and internal sensor. Zero calibration is recommended approximately every two weeks using clear water without ozone or color.
Product features (summary)- Test method: dual-path UV absorption for high measurement accuracy.
- Display units: mg/L or ppm selectable.
- Light pool system: separate light pool technology, leak-free and pressure-resistant; resistant to large-flow sampling impacts.
- Intelligent compensation: automatic zero adjustment and automatic light-source compensation; manual zero calibration supported.
- Display resolution and data: high-resolution display with selectable units shown on touch screen.
Technical parametersMeasuring ranges: 0-20 ppm; 0-50 ppm; 0-100 ppm; 0-120 ppm; 0-180 ppm; 0-200 ppm
Light source life: ~30,000-50,000 h
Sampling method: positive-pressure or negative-pressure sampling (use with pumps, valves, flow meters installed at the outlet)
Wetted materials: quartz, PTFE, PFA (sapphire, PTFE, PFA, FFPM)
Display interface: 4.3-inch high-definition touch screen
Display units: mg/L, ppm optional
Display resolution: 0.01 ppm
Sampling flow rate: 200-300 mL/min
Input pressure: < 0.1 MPa
Concentration error: max 0.3% FS
Linearity error: max 0.2% FS
Zero drift: < ±0.3% FS
Ambient temperature range: 0-50 °C
Interface: ferrule (PTFE) or optional 1/4 PFA interface
Sampling calibre: Φ6 (6 mm × 4 mm)
Communication mode: RS-485 (Modbus)
Output mode: 4-20 mA (active)
Power supply: 24 VDC
Dimensions: 150 mm (H) × 200 mm (W) × 80 mm (D)
Weight: ≈3 kg
Specifications- Model: OZ-300W UV (OZ-300W-HF for HF environments)
- Detection principle: UV dual-path absorption (Lambert-Beer law)
- Measuring ranges available: 0-20 ppm, 0-50 ppm, 0-100 ppm, 0-120 ppm, 0-180 ppm, 0-200 ppm
- Output: 4-20 mA; Communication: RS-485 (Modbus)
- Display: 4.3-inch HD touch screen; units mg/L or ppm
- Sampling flow: 200-300 mL/min; sampling calibre Φ6
- Wetted materials: PTFE, PFA, FFPM, quartz or sapphire
- Light source: UV LED 253.7 nm, life ~30,000-50,000 h, 3-year warranty for light source
- Accuracy: concentration error ≤0.3% FS; linearity error ≤0.2% FS; zero drift < ±0.3% FS
- Power: 24 VDC; Enclosure: splash-proof aluminum, IP65; Dimensions: 150×200×80 mm; Weight ≈3 kg