Product descriptionOPC-P5 Portable Mobile Oil Cleanliness Monitor is an instrument developed for oil contamination level detection in accordance with national and international standards (e.g. GB/T 18854-2002, ISO11171-1999). It is suitable for on-site and laboratory testing of hydraulic oil, lubricating oil, shale oil, transformer (insulating) oil, turbine oil, gear oil, engine oil, aviation kerosene, water-based hydraulic fluids, phosphate ester oils and other fluids. The unit can be installed on hydraulic transmission systems, oil filters, cleaning machines and test benches for online system oil cleanliness monitoring. It is used across aerospace, power generation, oil & gas, chemical, transportation, ports, metallurgy, machinery and automotive manufacturing sectors.
Key features- Detection principle: optical shading (light blocking) with high-precision laser sensor for particle counting.
- Designed for laboratory and on-site use; supports connection to an external pressure chamber for high-viscosity samples and degassing.
- Built-in data analysis displays particle-size data per channel and automatically determines sample cleanliness grade.
- Fluid path materials: 316L stainless steel and PTFE compatible with various organic solvents and oils.
- Volume-flush and time-flush modes for maintenance and cleaning.
- Integrated contamination standards: ISO4406, NAS1638, SAE4059 (integral & differential), GJB420A/B, ГОСТ17216, GB/T14039 and others.
- Calibration function compliant with GB/T21540, ISO4402, GB/T18854 and similar standards.
- Automatic data processing and printing; built-in water and temperature sensors.
- Bilingual interface (Chinese/English) with one-touch switching, presets, parameter editing and storage.
- Large internal storage and support for external USB storage; rugged embedded housing for portability.
Applications- On-site and laboratory oil cleanliness testing for hydraulic systems, lubricants and insulating oils.
- Online monitoring when installed on hydraulic transmissions, filters, cleaning machines and test benches.
- Applicable industries: aerospace, power, oil & gas, chemical, transport, ports, metallurgy, general machinery and automotive manufacturing.
Technical specifications- Model: OPC-P5
- Light source: semiconductor laser
- Offline detection flow: 5–60 mL/min (factory default 25 mL/min)
- Offline sample viscosity: ≤100 cSt (optional pressure chamber for higher viscosity)
- Particle size range: 1–500 μm
- Sensitivity: 0.8 μm or 4 μm (c)
- Coincidence error limit: 40,000 particles/mL
- Counting volume: 1–999 mL (10 or 20 mL recommended)
- Counting accuracy: error < ±10%
- Protection rating: IP67
- Test interval: 1 second – 24 hours
- Detection sample temperature: 0–80 °C
- Water activity reference: 0–1 aw
- Water content: 0–360 ppm
- Operating temperature: -20 – 60 °C
- Power supply: AC 220V ±10%, 50/60Hz or DC 12–40V
- Weight: 2.5 kg
- Dimensions (L×W×H): 275 × 220 × 107 mm