Product descriptionThe WGT-1200 Glove Integrity Tester is a wireless online multi‑unit instrument for detection of glove leaks, designed according to the positive‑pressure leak detection principle (GB/T 25915.7‑2010 / ISO 14644‑7:2004). It integrates a built‑in air pump and lithium battery, supports real‑time linking and centralized management of multiple detection units, and includes electronic signature and access control to meet 21 CFR Part 11 requirements. Replaceable glove trays support different glove specifications and sizes while tests are performed without removing the gloves. The device uses a precision pressure sensor to deliver fast, repeatable detection results and local flash memory to store test records.
Key features- Wireless data transmission to PC via Wi‑Fi for real‑time monitoring and record transfer.
- Centralized management software to control and monitor multiple detection units (up to 10 units).
- RFID identification or manual input for glove tracking; ID included in test records.
- Protection rating IP65 for disinfection and washdown procedures.
- Multiple preset test schemes to cover different glove materials and specifications.
- Pressure detection range 500–3000 Pa with 0.1 Pa resolution.
- Built‑in sanitary air pump for controlled inflation of gloves and cuffs.
- Automatic monitoring of sealing ring pressure and glove pressure during test cycles.
- Low, uniform air intake to prevent glove deformation during testing.
- Color touchscreen with intuitive interface and automatic program selection; date/time display.
- Portable design for easy placement and fast connection on site.
- History storage for more than 12,000 test records with transmission and query capability.
- Optional customizable glove trays/pallets for multi‑position operations (on demand).
- Designed to support FDA/GMP compliance and computer system validation.
Technical specifications- Model: WGT-1200
- Power supply: DC 12V, 20W
- Built‑in battery: 12V lithium battery
- Minimum detectable aperture: 300 µm
- Typical test duration: 3–8 min
- Pressure range: 500–3000 Pa
- Pressure resolution: 0.1 Pa
- Noise level: ≤50 dB(A)
- Communication: Wi‑Fi wireless network
- Gas source: Built‑in air pump (sanitary)
- Max sealing pressure: ≤200 kPa
- History capacity: >12,000 test sets (local flash storage)
- Glove recognition: RFID or manual input
- Weight: 4.5 kg
- Control units supported: up to 10 test units
- Control system: PC/software control (21 CFR Part 11 compatible)
Applications- Pharmaceutical industry: sterility and microbial limit inspections for purified water, water for injection and sterile preparations.
- Medical device industry: sterility and microbial limit testing for syringes, infusion sets, blood transfusion sets, IV catheters, etc.
- Food and beverage industry: integrity checks where glove/cuff seals are used.
- Environmental protection and other sectors requiring glove integrity verification.