Product DescriptionPackaging Sealing Integrity Tester MFT-1000 is designed to verify sealing integrity of pharmaceutical packaging and to prevent ingress of contaminants such as moisture, oxygen and microorganisms. It supports integrity testing for a wide range of pharmaceutical containers and spray cans using non‑destructive vacuum and pressure‑decay difference detection methods.
Application- Bottles, bags, boxes, ampoules, vials, cartridges, pre‑filled syringes (PFS), blow‑fill‑seal (BFS), form‑fill‑seal (FFS)
- Spray cans: plunger (piston) spray cans, bag‑lined spray cans, “energy jacket” spray cans, flexible tube spray cans
Features- Complies with USP <1207>, ASTM F2338 and FDA requirements
- Semi‑automatic detection suitable for small batches and multi‑variant testing
- Non‑destructive testing with high accuracy, repeatability and sensitivity
- Uses vacuum pressure and pressure‑decay difference detection methods
- Automatic conversion of leak rate into defect aperture (µm)
- Database storage of test results to support quality management
- Touch‑type human‑machine interface for simplified operation; manual sample loading/unloading after program selection
Advantage function- Autonomous optimization Linux system (ADPOT)
- Automatic flow measurement and aperture adjustment throughout the test process
- Automatic leak‑rate calibration function and supplied standard leaks (standard positive bottles)
- Four‑level user authority management supporting FDA 21 CFR Part 11
- Audit trail function with event logging
- Split design with test chamber above the host; multiple chambers available for different product types
- Support services: positive bottle production, standard leak verification, custom sample molds, methodological parameter development and verification
- Customizable test cavity to match product form for rapid and sensitive testing
Technical specifications- Model: MFT-1000
- Differential pressure sensor: ±2 kPa, error ≤ 0.5% F.S.
- Differential pressure resolution: 0.1 Pa
- Test pressure sensor accuracy: error ≤ ±1% F.S.
- Test pressure range: -0.1 to 0.2 MPa / -0.1 to 0.7 MPa / -0.1 to 1.0 MPa
- Minimum detectable aperture: 1.5 μm (5 μm for IV bag)
- Display: 10‑inch color touch screen
- Power: AC100V‑240V, 50/60Hz, ≤ 200W
- Weight: Host 30 kg
- Audit trail/history: ≥5 years storage, event logs queryable by time
- Authority management: username/password login, 4‑level authority
- Print function: built‑in printer; Data backup: USB export
- Signal output: 4‑20 mA, RS485, 12V alarm output
- Dimensions: 430 × 655 × 420 (Height1) × 330 (Height2) mm