Product OverviewZetron V4.0 Filter Integrity Tester is a compact, field-ready instrument for verifying integrity of filters and membranes (including ultrafiltration). The unit integrates an audit trail (3 years storage), user management and electronic signature functions to support regulated environments and traceable quality control.
Key Features- Supports all common integrity test methods: manual/basic/extended bubble point, pressure-hold, diffusion flow, water intrusion and ultrafiltration membrane tests.
- Authority management with username/password, role-based permissions and electronic signature compliant with FDA 21 CFR Part 11.
- Up to 120 reservation solution groups for repeatable, preconfigured test setups.
- Rich connectivity: standard RS232 and USB; customizable industrial bus and control ports for system integration (e.g., online sterilization control).
- Data handling: built-in printer, USB export including test curves and storage of >15,000 record sets.
- Independent R&D support for customized system designs and technical integration.
Technical Specifications- Model: V4.0
- Power supply: 100–260 VAC, 50 Hz, 120 W
- Operation pressure: 100–10000 mbar (1.5–145 psi)
- Units: mbar
- Operating conditions: +5 ℃ ~ +40 ℃; Relative humidity 10–80%
- Dimensions (mm): 400 (Depth) × 240 (Width) × 270 (Height1) × 220 (Height2)
- Test accuracy: Upstream volume ≤ 4%; Bubble Point ≤ 50 mbar; Diffusion Flow ≤ 4%; Water Intrusion ≤ 0.01 ml
- Test ranges: BP 100–8000 mbar; DF 1–600 ml/min; WI 0.01–100 ml/min
- Applied range: symmetric/asymmetric membranes, needle filters, filter bags, flat filters, large-volume cartridges, ultrafiltration membranes/columns and irregular filters
- Reservation solutions: 120 sets
- Audit trail retention: 3 years
- Display: 7-inch color touchscreen
- Serial connection: RS232 / USB
- Languages: Chinese / English
- Work mode: Online / Offline
- Other: 1D/2D code support, barcode scanner, optional lithium battery
- Applicable environment: Above D level
- Weight: 8 KG