Product overviewIntegtest Filter Integrity Tester V8.0 is designed to test and verify filter integrity and performance in industrial environments where removal of impurities and microorganisms from fluids is critical. The instrument supports online and offline modes, bilingual interface (Chinese/English) and expandable communication interfaces.
Key features- Multiple integrity test methods: bubble point, pressure-hold, diffusion flow, water immersion and ultrafiltration tests.
- Optimized Linux operating system and automatic self-test at power-on.
- 10-inch true-color touchscreen HMI for intuitive operation.
- High-accuracy pressure sensors; selectable pressure units (mbar, kPa, psi, kgf/cm2).
- Up to 1000 pre-stored programs; supports testing of most cartridge filters up to 12 cores and 20 inches.
- User management: up to 1000 accounts, 4 authority levels, password login and electronic signature; built-in audit trail with exportable, irreversible records (≥5 years).
- Data interfaces: RS232, USB, PC printing control, USB export of records and curves; wireless communication supported; built-in thermal printer.
- Storage capacity up to 300,000 records; configurable test programs and expandable network interfaces.
- Compact: approx. 10 kg; protection rating IP54.
Applications- Pharmaceutical industry: sterile and microbial-limit testing of purified water, water for injection and sterile preparations.
- Medical device manufacturing: sterility checks for water, syringes, infusion/transfusion sets, IV catheters.
- Food & beverage production and environmental protection applications.
ParametersModel: V8.0
Power supply: 100–240 VAC, 50 Hz, 120 W; operational spare battery
Operating pressure: 100–10000 mbar (max 150 psi)
Pressure units: mbar, kPa, psi, kgf/cm2
Operating conditions: +5°C to +40°C; RH 10–80%
Display: 10-inch HD color touchscreen
Dimensions (mm): 480 × 300 × 225
Weight: ≈10 kg
Work mode: Online / Offline
Protection: IP54
Record storage: up to 300,000 sets; audit trail retention ≥5 years
Characteristics / Technical specifications- Test functions: manual/basic/extended bubble point, pressure hold, diffusion flow, water immersion, ultrafiltration membrane tests.
- Test ranges & accuracy: BP 100–8000 mbar (±50 mbar); DF 1–900 ml/min (±4%); WI 0.01–100 ml/min (±0.01 ml); upstream volume ±4%.
- User management & audit: 1000 user accounts, 4 authority levels, electronic signature, irreversible audit trail.
- Interfaces & communication: RS232, USB, PC printing control, USB export, wireless support; expandable network interfaces.
- Storage & programs: 1000 pre-stored programs; historical storage ~300,000 records.
- Operating system: optimized Linux.