Product overviewThe wireless glove integrity tester WGT-1200 is designed according to the positive-pressure leak detector principle recommended by GB/T 25915.7-2010/ISO 14644-7:2004. It is a dedicated instrument for detecting glove leaks, combining WiFi wireless communication, an integrated sanitary air pump and a 12V lithium battery for autonomous operation and real-time linkage of multiple test units.
Key features- WiFi wireless communication for real-time data transmission to PC and central management software.
- Upper‑computer management software supports concurrent supervision of multiple detection units.
- RFID chip identification and option for manual glove ID input; glove ID included in test records.
- IP65 protection for easy disinfection and equipment cleaning.
- Multiple preset test schemes for different glove specifications and materials.
- Pressure detection range 500–3000 Pa, pressure resolution 0.1 Pa.
- Built-in sanitary air pump for controlled inflation of gloves and cuffs; minimal, frequent air intake to avoid glove deformation.
- Automatic monitoring of inflatable sealing ring and glove pressure during tests.
- Color touch screen with intuitive interface; automatic program selection and date/time display.
- Portable design, easy to connect and deploy; supports fast connections and mobility.
- Onboard FLASH memory and history storage (capacity >12,000 test sets) with real-time transmission capability.
- Customizable, replaceable glove trays to support different glove sizes and multi-pallet operations.
- Designed to comply with FDA/GMP requirements and computer system verification; software compliant with 21 CFR Part 11 for electronic records and signatures.
Applications- Pharmaceutical industry: sterile inspection and microbial limit testing for purified water, water for injection and sterile preparations.
- Medical device industry: sterility and microbial limit inspection for syringes, infusion sets, blood transfusion sets, IV catheters and related accessories.
- Food and beverage industry for integrity checks where applicable.
- Environmental protection and other sectors requiring glove leak and integrity testing.
Technical parameters (table)Power supply: DC 12V 20W
Built-in battery: 12V lithium battery
Test accuracy: minimum detectable aperture 300 μm
Typical test duration: 3–6 min
Pressure range: 500–3000 Pa
Pressure resolution: 0.1 Pa
Noise: ≤50 dB(A)
Communication method: WiFi wireless network
Gas source: built-in sanitary air pump
Sealing pressure capacity: ≤200 kPa
History record capacity: 12,000 sets
Glove recognition: RFID / manual input
Weight: 4 kg
Online glove tray: customizable replacement
Control units configurable: up to 10 test units controllable
Control system: computer control
Characteristics / technical specifications- Model: WGT-1200
- Power supply: DC 12V, 20W
- Built-in battery: 12V lithium battery
- Minimum detectable aperture: 300 μm
- Typical test duration: 3–6 minutes
- Pressure detection range: 500–3000 Pa
- Pressure resolution: 0.1 Pa
- Noise level: ≤50 dB(A)
- Communication: WiFi wireless network
- Gas source: built-in sanitary air pump
- Maximum sealing pressure: ≤200 kPa
- History storage: up to 12,000 test sets
- Glove recognition method: RFID or manual input
- Weight: 4 kg
- Supports control of up to 10 test units via PC management software