Product overviewThe Zetron BGT-120 bag integrity tester is a portable pressure-decay instrument for leak detection in sterilization solution bags, disposable solution bags, transfer bags and respirators. It features a small-frequent air intake mode, proven pressure-control technology, access control and electronic signature support, a built-in printer and a color touchscreen that can automatically select test programs and display date/time. The unit operates without additional equipment and stores test history locally.
Application- Integrity testing of sterilization solution bags, disposable solution bags, transfer bags and respirators to verify sterility requirements after cleaning and use.
- On-site routine quality control, batch release and maintenance/validation workflows (IQ/OQ/PQ/SOP).
Standards & Compliance- Pressure-decay test method recognized by ASTM F2095-01.
- Complies with GAMP5 guidance and FDA 21 CFR Part 11 for electronic records and electronic signatures.
- Supports IQ, OQ, PQ and SOP client-side authentication and validation processes.
Key features- Proven pressure-control technology for high testing accuracy.
- Small, frequent air intake mode ensures uniform intake speed and prevents bag deformation.
- Access control and electronic signature design meeting FDA 21 CFR Part 11 requirements.
- Built-in printer; ability to print, store and retrieve test records; historical storage ≥15000 groups.
- Color touchscreen with intuitive interface and automatic test-program matching.
- Portable, quick to connect and easy to move; built-in lithium-ion battery option.
- Supports IQ, OQ, PQ and SOP validation workflows.
Main parametersWorking power supply: AC 220V 50Hz; Built-in battery: Lithium-ion; Rated power: 120W; Test duration: 2–8 min; Pressure range: 0–5000 Pa; Response value: ≤50 Pa; Noise: ≤50 dB(A); Operation mode: Continuous; External gas source: Compressed air, nitrogen, inert gas; Historical records: ≥15000 groups; Accuracy: 0.1 Pa; External dimensions (W×D×H): 420×330×180 mm; Weight: 5 kg; Audit-trail retention: ≥3 years.
Characteristics / technical specifications- Model: BGT-120
- Brand: Zetron
- Working power supply: AC 220V 50Hz
- Built-in battery: Lithium-ion batteries
- Rated power: 120W
- Test duration: 2–8 minutes
- Pressure range: 0–5000 Pa
- Response value: ≤50 Pa
- Noise: ≤50 dB(A)
- Operation mode: Continuous
- External gas source: Compressed air, nitrogen, inert gas
- Historical records storage: ≥15000 groups
- Accuracy: 0.1 Pa
- External dimensions (W×D×H): 420×330×180 mm
- Weight: 5 kg