OverviewTA-1.0 Total Organic Carbon Analyzer is a high-precision offline TOC instrument designed for laboratory testing of pharmaceutical waters and ultrapure water. It combines conductivity-difference detection with UV catalytic oxidation to provide accurate and rapid TOC measurements while supporting regulatory testing workflows.
Operation PrincipleOrganic compounds are oxidized by a UV lamp and converted to carbon dioxide (CO2). Detection is performed by direct conductivity measurement. TOC is calculated as the difference between total carbon (TC) after oxidation and total inorganic carbon (TIC): TOC = TC - TIC.
Main features- No need for nitrogen, oxygen, acidic reagents or external oxidants, reducing consumable requirements.
- High automation level: supports automatic sampler and automated system adaptability validation.
- Limit alarm design: automatic alarms when sample results exceed preset thresholds.
- Audit-ready functions and user access control to support regulatory compliance.
ParametersPower supply: 100-260VAC 50Hz/60Hz, 120W
Conductivity detection range: 0.055 µs/cm - 8.000 µs/cm
Measurement range (TOC): 0.001 mg/L ~ 1 mg/L (1 ~ 1000 ppb)
Accuracy: ±3%
RSD (Repeatability): ≤ 3%
Resolution: 0.001 mg/L
Response time: 5 mins
Measuring limit: 0.001 mg/L
Sample temperature: 1-99 ℃
Applied range: Offline laboratory testing, cleaning validation
Audit trail, records: ≥ 3 years storage
Authority management: Username and password login to meet FDA 21 CFR Part 11 requirements
Print function: External printer
Data copy: USB disk export
Display: Color touchscreen
Weight: 8.5 kg
Dimensions (mm): 400 × 240 × 270
ApplicationSuitable for offline laboratory testing and monitoring of pharmaceutical water (purified water, water for injection), cleaning validation, and applicable to electronics and food industries where TOC control is required.
Technical specifications- Model: TA-1.0 Offline
- Detection technology: UV catalytic oxidation + direct conductivity measurement
- TOC measurement range: 0.001–1 mg/L (1–1000 ppb)
- Resolution: 0.001 mg/L
- Accuracy: ±3%
- Repeatability (RSD): ≤3%
- Response time: 5 minutes
- Power: 100-260VAC, 50/60Hz, 120W
- Display: Color touchscreen
- Data storage: ≥3 years (audit trail)
- Data export: USB disk
- Compliance: Designed to meet national pharmacopoeia test schemes and support FDA 21 CFR Part 11 authority management
- Weight & Dimensions: 8.5 kg; 400 × 240 × 270 mm