Product overviewThe TA-201E Total Organic Carbon Analyser measures TOC by high-temperature catalytic combustion and sequential NDIR CO₂ detection. Samples and purification gas (high‑purity oxygen) are routed to a high‑temperature combustion tube and a low‑temperature reaction tube to produce CO₂ from organic and inorganic carbon. CO₂ from both paths is measured by NDIR; TC and IC are obtained separately and TOC is calculated as TOC = TC − IC.
Features- High‑temperature catalytic combustion efficiently oxidizes refractory organic carbon, enabling analysis of high‑concentration TOC samples.
- Fast cycle time: typical analysis 1–4 minutes.
- Multiple safety protections on the combustion furnace and independent overheat protection with automatic shutdown.
- Real‑time flow monitoring to ensure stable carrier flow and reliable results.
- Bidirectional pipeline cleaning and purging; internal circuits accessible for cleaning to reduce downtime.
- Automatic waste discharge and automated acid dosing with controlled inlet volume.
- Low sample and reagent consumption: per measurement ~0.5 μL high‑purity water, 2 mL acid reagent (IC), and ≈2000 mL O₂ (standard conditions: 100 mL/min, purge 20 min).
- NDIR CO₂ detection with good linearity; internal data processor converts CO₂ signal to peaks and computes TOC.
- Combustion at ~680 ℃ (extendable to 1200 ℃) below melting point of most salts, prolonging catalyst and combustion tube life—suitable for saline samples.
- 7‑inch high‑resolution touch screen and intelligent UI for straightforward operation.
Functional parametersModel: TA-201E
Measurement range: 0–1000 mg/L (non‑diluted); up to 0–30000 mg/L with dilution
Repeatability: ≤3%
Indication error: TC ±0.1% F.S or ±5% (larger); IC ±0.1% F.S or ±4% (larger)
Linearity: R² ≥ 99.9%
Limit of detection: 0.5 mg/L
Analysis time: 2–4 min
Injection volume: 10 μL–500 μL
External storage: USB U‑disk
Power supply: AC 220V ±10% 50/60Hz (reliable grounding required)
Power consumption (heating): 1000 W
Operating ambient: 0–40 ℃, relative humidity 10–85%
Installation: indoor, allow ≥20 cm clearance for heat dissipation
Gas: high‑purity oxygen (99.999%)
High‑purity water: CO₂‑free water (TOC ≤ 0.5 mg/L)
Acid reagent: 10% phosphoric acid
Combustion temperature: 680 ℃–1200 ℃
ApplicationsDetermination of TOC in surface water, groundwater, domestic sewage and industrial wastewater. Applicable to environmental monitoring, urban water supply and drainage, public health control, chemical industry and similar sectors requiring reliable TOC measurement.
Technical specifications- Model: TA-201E
- Measurement principle: high‑temperature catalytic combustion + NDIR CO₂ detection
- Range: 0–1000 mg/L (non‑diluted); up to 0–30000 mg/L with dilution
- Repeatability: ≤3%
- Detection limit: 0.5 mg/L
- Typical analysis time: 2–4 minutes
- Injection volume: 10–500 μL
- Power: AC 220V ±10% 50/60Hz; heating power 1000 W
- Gas requirement: O₂ 99.999%
- Acid reagent: 10% H₃PO₄
- Operating environment: 0–40 ℃, 10–85% RH