Product overviewThe LPC-16DA Intelligent Particle Detector is designed for quantitative and sizing analysis of particulate contamination in pharmaceutical liquids and related matrices. It integrates an imported laser light source with a compensation circuit, a high-pressure syringe pump sampling system and 316L stainless steel/PTFE flow path materials to allow direct testing of aqueous, organic solvent and oil-based samples.
Features- Pharmacopoeia-compliant: meets relevant 2020 editions of the Chinese Pharmacopoeia and the United States Pharmacopeia for particle content and size testing of injections, sterile powders, infusion equipment and packaging materials.
- Laser detection with compensation circuit supports accurate measurement of colourless and coloured samples, including electrolyte-free samples.
- Interchangeable high-pressure syringe pump sampling with adjustable injection volume to accommodate high-viscosity products and altitude variations.
- Injection slit and flow path constructed from imported 316L stainless steel and PTFE for direct detection of organic solvents and oil matrices.
- Dedicated programs for large-volume infusion tests and small-volume injections with automatic conversion of results to particle size and number per container as required by pharmacopoeias.
- Special test routines for precision filters, anesthesia packs and filtration-rate verification for infusion equipment.
- Large colour touch display running Android for intuitive operation and simultaneous multi-channel data display.
- Data management: support for up to 1000 sample names, 1000 user accounts and storage for up to 1,000,000 records; reserved computer audit interface, automatic data processing and multiple print modes; Android system backup.
ApplicationsIntended for pharmaceutical manufacturers, medical device makers, packaging suppliers and testing laboratories. Typical uses include particle contamination testing of injections, sterile powders and small-volume injections; filtration-rate verification of infusion equipment (e.g., anesthesia bags, precision filters); and compliance testing of packaging materials (glass vials, rubber stoppers). Suitable for laboratory R&D, GMP production quality control and third-party testing agencies requiring high-precision data.
Technical specifications- Model: LPC-16DA
- Test range: 1–500 μm
- Channel settings: 16 channels (examples: ≥2 μm, ≥3 μm, ≥4 μm, ≥5 μm, ≥6 μm, ≥8 μm, ≥10 μm, ≥12 μm, ≥15 μm, ≥19 μm, ≥20 μm, ≥21 μm, ≥25 μm, ≥50 μm, 75 μm, ≥100 μm)
- Injection volume: 5 ml ± 3%; arbitrary sampling 0.2–99,999 ml
- Counting range: 0–9,999,999 particles
- Relative standard deviation (RSD): ≤ 2% (standard particles 1000–1500 particles/ml)
- Accuracy: specified value ±20% (pharmacopoeia standard)
- Channel resolution: ≥ 68% (pharmacopoeia standard)
- Working temperature: 10–35 ℃
- Power supply: AC 220V ± 10%; 50Hz; ≤60W
- Weight: 22 kg
- Dimensions: 440 × 355 × 475 mm