Product overviewThe OPC-P50 is a portable hydraulic oil cleanliness analyzer designed for laboratory, on-site and inline monitoring of oil contamination levels in accordance with national and international standards (e.g. GB/T 18854-2002, ISO11171). It supports a wide range of fluids including hydraulic oils, lubricating oils, transformer oil, turbine oil, gear oil, engine oil, aviation kerosene, water-based hydraulic fluids and phosphate ester oils.
Key features- Shading (light blocking) measurement principle with high-precision semiconductor laser sensor for accurate particle counting.
- Supports offline (5–60 mL/min) and online measurement; optional pressure-reducing device enables high-pressure online monitoring up to application-specific limits.
- External pressure chamber connectivity for positive/negative pressure and degassing of high-viscosity samples.
- 316L and PTFE wetted flow path materials for compatibility with a wide range of solvents and oils.
- Built-in data analysis with per-channel particle size display and automatic sample grading; supports standards ISO4406, NAS1638, SAE4059, GJB420A/B, GB/T14039, ГОСТ and others.
- Calibration support for GB/T21540, ISO4402 and GB/T18854; automatic data processing and printing functions.
- Integrated water and temperature sensors; bilingual Chinese/English UI with one-touch switching; 10.2-inch display.
- Robust IP67 enclosure, internal storage for 1,000 records and USB external storage support; rechargeable battery for field use.
Applications- Inline and offline oil cleanliness monitoring for hydraulic systems, filters, cleaning units and test benches.
- Used across aerospace, power generation, oil & gas, petrochemical, transport, ports, metallurgy, heavy machinery and automotive manufacturing.
Technical specifications- Model: OPC-P50
- Light source: semiconductor laser
- Particle size range: 1–500 μm
- Offline flow rate: 5–60 mL/min
- Offline sample viscosity: ≤100 cSt (pressure chamber optional for higher viscosity)
- Online detection pressure: 0.1–0.6 MPa (optional pressure reducing device for higher pressures)
- Sensitivity: 0.8 μm or 4 μm (configurable)
- Counting volume: 1–999 mL; counting accuracy: ±10% max
- Data storage: 1,000 records internal + USB support
- Protection: IP67
- Detection temperature range: 0–80 °C; operating temperature: -20–60 °C
- Water content range: 0–360 ppm (±10%); water activity: 0–1 aw (±0.05 aw)
- Power: DC 12.6V / AC 100–240V; battery: 8000 mAh
- Weight: 5.5 kg; dimensions: 365 × 295 × 170 mm; screen: 10.2 inches