ProductIntroductionThe Glove Integrity Tester GT-2.0 Online is designed according to the positive-pressure leak detection principle referenced in GB/T 25915.7-2010 and ISO 14644-7:2004. It is a dedicated instrument for detecting leaks in gloves used in controlled environments. The unit requires no additional equipment, does not require removal of gloves during testing, and provides fast, repeatable leak detection.
Features- Small-volume, high-frequency air inlet mode with uniform inlet velocity to avoid glove deformation during testing.
- Color touchscreen and intuitive interface; automatic parameter selection and matching; date and time display.
- Authority management and electronic-signature capability compliant with 21 CFR Part 11 for regulated environments.
- Portable design for easy transport and rapid connection; historical test records can be printed, stored and queried.
Technical parameters- Power supply: AC 220 V 50 Hz
- Rated power: 120 W
- Test duration: 2–8 min (configurable)
- Pressure range: 500–5000 Pa
- Response value (sensitivity): ≤ 50 Pa
- Noise level: ≤ 50 dB(A)
- Operating mode: continuous
- External gas source: compressed air, nitrogen, inert gas
- Maximum sealing pressure: ≤ 0.8 MPa
- Other functions: audit trail, USB data export (optional)
- Historical records: 10,000 groups
- Print function: built-in printer
- Overall dimensions: 400 x 260 x 270/250 mm
- Weight: 7.5 kg
- On-line gloves tray: custom-made
- Audit-trail storage (optional): ≥ 3 years
Applications / NotesInstrument intended for glove leak testing in cleanrooms and other controlled environments where documented verification and audit trail are required. Suitable for use in pharmaceutical, biotechnical and other regulated production or testing areas.